24
TIME BETWEEN SAMPLES
Allows the operator to input the desired time between samples.
When TIME BETWEEN SAMPLES is selected and you press
ENTER two choices will presented for your selection.
EQUAL INTERVALS
Allows the operator to input the desired time between samples
in an hours:minutes format. All time intervals will be the same
or equal. You can select any time period between 0 minute to
99 hours 59 minutes. The time for this option should be set
to a value greater than 0 except for applications where the
number the # CONTACT CLOSURES is chosen to control time
between samples. The display appears as follows:
NOTE: If your are using the # CONTACT CLOSURES menu
option the time in this option should be set to 00:00 HR:MN,
otherwise the two features will interact. This will cause the
timer to enact a sampling cycle rather than the set number of
contact closures if the time interval occurs before the # OF
CONTACT CLOSURES..
UNEQUAL INTERVALS
Allows the operator to input desired time between samples
in an hours:minutes format where the time between samples
can vary. The operator is allowed a maximum of 12 different
time periods between samples. If more than 12 samples were
selected in NUMBER OF SAMPLES menu option the remaining
time periods will repeat time period #12.
If only 5 samples were selected in NUMBER OF SAMPLES
only 5 time periods will appear in the selection menu. The
display appears as follows:
HINT: the display is only 2 lines, so you will only see 2
choices at one time. Remember the arrow in the first column
indicates whether there are choices above or below the ones
you are viewing.
Once you have made your selections and pressed enter the
display changes to read:
Use the down arrow key to navigate and select # OF CONTACT
CLOSURES and press ENTER. The display changes to read:
# OF CONTACT CLOSURES
This menu option will not be visible if you selected REMOTE &
DELAYED START in the PROGRAM START option and selected
the MAINTAINED CONTACT menu option. The system will be
started by a single contact closure. The system will be take a
sample after the specified number of contact closures occurs.
The number of contact closures available is 0-9999. The
default is 0.
Once you have made your selections and pressed enter the
display changes to read:
Use the down arrow key to navigate and select DEFAULT
PROG SETTINGS and press ENTER.
g
EQUAL INTERVALS
c
UNEQUAL INTERVALS
EQUAL INTERVALS
00:15 HR:MN (0-99:59)
g
#1 00:15 HR:MN
c
#2 00:30 HR:MN
c
#2 00:45 HR:MN
c
#3 01:00 HR:MN
c
#11 02:45 HR:MN
g
#12 03:00 HR:MN
g
TIME BETWEEN SAMPLES
c
# OF CONTACT CLOSURES
# OF CONTACT CLOSURES
(0-9999)
g
TIME BETWEEN SAMPLES
c
DEFAULT PROG SETTINGS