9-6
Radical-7R Pulse CO-Oximeter Operator’s Manual
9
S e r v i c e / M a i n t e n a n c e
6.
Connect a patient cable and sensor and verify that the voltage on pins 9 and 15 are
between 0V and 1.0V while measuring a saturation and pulse rate.
Battery Test:
1.
Fully charge the Radical-7R by placing the Handheld into the Docking Station and
connecting the AC power.
2.
Verify that the green Handheld Battery Indicator LED is illuminated.
3.
When the Radical-7R is fully charged the green Handheld Battery Indicator turns off.
4.
Turn the Radical-7R on and verify that the Battery indicator shows a full charge.
Service and Repair
REPAIR POLICY
Masimo or an authorized Service Department must perform warranty repair and service. Do not
use malfunctioning equipment. Have the instrument repaired.
WARNING:
DO NOT REMOVE THE COVER OF THE MONITOR EXCEPT FOR
BATTERY REPLACEMENT. AN OPERATOR MAY ONLY PERFORM
MAINTENANCE PROCEDURES SPECIFICALLY DESCRIBED IN THIS
MANUAL. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL
TRAINED IN THE REPAIR OF THIS EQUIPMENT.
Please clean contaminated and/or dirty equipment before returning, following the cleaning
procedure described in Section 9,
Cleaning
. Make sure the equipment is fully dry before packing.
To return the Radical-7R instrument for service, please follow the Return Procedure.
RETURN PROCEDURE
Please clean contaminated/dirty equipment before returning. Make sure the equipment is fully
dry before packing. Call Masimo at 800-326-4890 and ask for Technical Support. Ask for an RMA
number. Package the equipment securely – in the original shipping container if possible – and
enclose or include the following information and items:
■
A letter describing in detail any difficulties experienced with the Pulse CO-Oximeter.
Please include the RMA number in the letter.
■
Warranty information – a copy of the invoice or other applicable documentation must be included.
■
Purchase order number to cover repair if the Pulse CO-Oximeter is not under warranty, or
for tracking purposes if it is.
■
Ship-to and bill-to information.
■
Person (name, telephone/Telex/fax number, and country) to contact for any questions
about the repairs.
■
A certificate stating the Pulse CO-Oximeter has been decontaminated for bloodborne pathogens.