Radical-7R Pulse CO-Oximeter Operator’s Manual
9-3
9
CONFIGURATION #1:
Radical-7R Pulse CO-Oximeter configured to only include the Handheld battery (standard
configuration); the Docking Station battery is excluded.
NOTE:
For this configuration, it is advisable to operate only the Radical-7R Handheld instrument
when running on battery power. Although it is possible to operate the entire Standalone
instrument (the Handheld attached to the Docking Station, with the Handheld battery powering
the Docking Station as well) on battery power, the capacity of the Handheld battery pack is
insufficient to support this mode for long periods of time.
RADICAL-7R CONFIGURATION
MINIMUM RUN-TIME
HANDHELD ONLY
1 hr
CONFIGURATION #2:
Radical-7R configured to include the Handheld and the Docking Station battery:
RADICAL-7R CONFIGURATION
MINIMUM RUN-TIME
HANDHELD AND DOCKING STATION
6 hr
REPLACING THE BATTERIES
Before installing or removing the battery, make sure the AC power cord is removed and power to
the Pulse CO-Oximeter is turned off.
To replace the Handheld battery, follow these instructions:
1.
Turn the Radical-7R Pulse CO-Oximeter Handheld off and remove the patient cable
connection. Detach the Radical-7R Pulse CO-Oximeter Handheld from the Docking Station
(if docked).
2.
Loosen the closure screw on the battery compartment door and lift out the battery.
3.
Take a new battery, and place it in the compartment.
4.
Tighten the closure screw.
5.
Place Handheld into Docking Station, turn on line power and charge battery according to
this Section, Battery Operation and Maintenance.
CAUTION:
FOLLOW LOCAL GOVERNING GUIDELINES FOR PROPER DISPOSAL OF
INTERNAL BATTERIES. DO NOT INCINERATE
.
WARNING:
THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/
OR
REMOVED
FROM
DOCKING
STATION
BY
QUALIFIED
PERSONNEL ONLY.
REPLACING THE FUSES
Should a power problem blow one or both of the fuses in the power entry module on the rear
panel, the fuse(s) will need to be replaced.
S e r v i c e / M a i n t e n a n c e
B a t t e r y O p e r a t i o n a n d M a i n t e n a n c e
( c o n t i n u e d )