Radical-7R Pulse CO-Oximeter Operator’s Manual
8
S e n s o r s & P a t i e n t C a b l e s
•
Newborn
Neonatal
•
Trauma
M a s i m o S e n s o r s c o n t i n u e d
SENSOR ACCURACY
See Sensor Directions for Use (DFU) for sensor accuracy specifications for: SpO2, SpMet,
SpCO, SpHb, pulse rate and respiration rate.
CLEANING AND REUSE OF MASIMO REUSABLE SENSORS AND CABLES
Reusable sensors and patient cables can be cleaned per the following procedure:
1.
Remove the sensor from the patient.
2.
Disconnect the sensor from the patient cable.
3.
Disconnect the patient cable from the monitor.
4.
Wipe the entire sensor and/or patient cable clean with a 70% isopropyl alcohol pad.
5.
Allow to air dry thoroughly before returning it to operation.
CAUTION:
CAREFULLY ROUTE PATIENT CABLES TO REDUCE THE POSSIBILITY OF
PATIENT ENTANGLEMENT OR STRANGULATION.
REATTACHMENT OF A SINGLE USE ACOUSTIC RESPIRATION SENSOR
■
A single use Acoustic Respiration sensor may be reapplied to the same patient if it is
dry, clean, free of debris and oil and the adhesive still adheres to the skin.
REATTACHMENT OF SINGLE USE ADHESIVE SENSORS
Single use sensors may be reapplied to the same patient if the emitter and detector windows are
clear and the adhesive still adheres to the skin.
NOTE:
If the sensor fails to track the pulse consistently, the sensors may be incorrectly
positioned. Reposition the sensor or choose a different monitoring site.
CAUTION:
DO NOT ATTEMPT TO REPROCESS, RECONDITION OR RECYCLE ANY
MASIMO SENSORS OR PATIENT CABLES AS THESE PROCESSES MAY
DAMAGE THE ELECTRICAL COMPONENTS, POTENTIALLY LEADING TO
PATIENT HARM.
CAUTION:
TO PREVENT DAMAGE, DO NOT SOAK OR IMMERSE THE SENSOR IN ANY
LIQUID SOLUTION. DO NOT ATTEMPT TO STERILIZE BY IRRADIATION, STEAM,
AUTOCLAVE OR ANY METHOD OTHER THAN ETHYLENE OXIDE AS INDICATED.
WARNING: TO AVOID CROSS CONTAMINATION ONLY USE MASIMO SINGLE
USE
SENSORS ON THE SAME PATIENT.