Model: TC5396
masibus
Doc. Ref. no. : - m105/om/101
Issue no. 00
User’s Manual
Page 7 of 40
2.4
Power Supply
Standard
85-265VAC/ 100-300VDC
Optional
18-36VDC
Power consumption
<10 VA (Aprrox)
Isolation (Withstanding voltage)
Between primary terminals* and secondary terminals**: At least 1500 V AC for 1 minute
Between primary terminals* and grounding terminal: At least 1500 V AC for 1 minute
Between grounding terminal and secondary terminals**: At least 1500 V AC for 1 minute
Between secondary terminals**: At least 500 V AC for 1 minute
* Primary terminals indicate power terminals and relay output terminals.
** Secondary terminals indicate analog I/O signal and Communication O/P.
Insulation resistance: 20MΩ or more at
500 V DC
2.5
Environmental
TEMPCO
For Input to PV Display < 100ppm. FOR Display to
Retransmission and Control output < 100ppm
Humidity
30% to 95% RH (Non-Condensing)
Ambient temperature
0 to 55°C
Storage Temperature
0
to 80°C
2.6
Physical
Mounting Type
Panel
Size H x W x D
100 mm x 100 mm x 55 mm
Front Bezel
100 mm x 100 mm
Panel Cutout
92 mm x 92 mm
Depth
Behind
The
Panel
52 mm
Weight
300g Approx.
Enclosure Material
Front: Polycarbonate, Base: ABS
Enclosure Protection
IP 20
Terminal & Cable Size
Barrier type terminal, cable 2.5 mm
2
Advance Feature
Input scalability for linear input
Three years calibration with Auto Zero and Auto Span
Average Energy Demand parameter as a diagnostic feature