Min & Max
Shows the Minimum (in place of mV
0
) and
Maximum value both together. This option
available only for only TC Display mode.
Available Options for TC Source Mode:
Options
Icon Description
None
-
No info is visible.
Actual Value
Shows the Actual Thermocouple
Temperature/mV value without any scaling.
This option is available only if TC Display
Mode is Percentage.
mV
Shows the Thermovoltageaccording to
Temperature including CJ temperature mV.
The mV which is sourced through TC
Terminal.
Reading
Feedback
Shows the Feedback Temperature/mV
Reading.
When PCE-MCA 50 generate mV, it uses its
own measurement function to control the
generated value. This feedback
measurement is shows if this option is
selected.
%Error
Shows the error in % between the Desired
Source Value and Feedback value.
mV Feedback
Shows the Feedback mV Reading.
mV w/o CJC
Shows the Thermovoltage according to
Temperature (for CJ Temperature = 0 °C)
Filter(sec)
1
st
Order IIR Low Pass Filter for TC Measure Reading.
This option is available only for TC mode is Measure.
Filter is useful when a measurement signal contains unwanted noise.
Range:
0.0 to 60.0 sec