Parameter Name Description / Options
Low
Starting Value of Step.
Enter value according to RTD Display Mode. If display mode is
actual enter value in ohms/°C and if display mode is % enter value
in %.
High
Ending Value of Step.
Enter value according to RTD Display Mode. If display mode is
actual enter value in ohms/°C and if display mode is % enter value
in %.
Manual
(Output Type)
Step Manual Mode Selection CheckBox.
Ticking this checkbox will enable Step Manual Mode. And Un-
ticking will enable Auto Step Mode.
Step Time (s)
Enter the time for a single step in seconds,
Range:
1 to 9999
This parameter is enabled only for Auto Step Mode (Manual
CheckBox is
Un-Checked)
Step Definition
Step Definition for the Step function.
Available Options:
Step Value
User Defined
Step
Step Value in Temperature/Ohms/% according to RTD Display
Mode and RTD unit.
Only appears if Step Definition is Step Value.
Define Steps
User Defined Step value for Manual and Auto Step Mode.
This option appear only if Step Definition is User Defined.
Maximum 10 step value can be configured. First enter the no of
step and then define step value in serial order.