Model: LC5296-AT /LC5248E-AT/LC5296V-AT 

 

masibus

 

Doc. Ref. no. : - m61C/om/301 
Issue no. 13 

 
User’s Manual 

 

Page 36 of 44 

 

9.  CALIBRATION PROCEDURE 

9.1 Procedure for CAL-zero and CAL-span 

The instrument is factory calibrated for the specified range, but due to long term drift of 
components,  re-calibration  may  be  necessary  in  some  cases.  For  calibrating  the 
instrument  a  reliable  source  is  required.  This  source  should  be  at  least  ten  times 
accurate compared to the range of the instrument. 
 
The unit can be calibrated without opening it and without trim pots.  
 
For Entering into the Calibration Mode, Please refer Menu Layout. 

After  applying  appropriate  Input  from  the  calibrator  source,  press  ‘INCREMENT’  OR 
‘DECREMENT KEY’ to bring the actual process value on display. 

 

 

Example:- 
At zero calibration reading expected on the display is 100 and it shows 107, adjust the 
process value to 100 by using 

‘DECREMENT KEY’. Now press ‘SET’ to store the calibration 

parameter  in  non-volatile  memory.  Similarly  one  can  calibrate  Ambient,  SPAN  and 
retransmission parameters. 
 
For  calibrating  i/p,  both  zero  and  span  are  calibrated.  Here  one-shot  calibration 
technique is used, i.e. the zero and span are calibrated once. Individual zero or span can 
also  be  calibrated;  first  calibrate  zero  and  then  span.  While  calibrating  thermocouple 
type  input,  first  calibrate  the  ambient  (if  required)  and  then  continue  with  the  span 
calibration of the same.  

 

9.2 Procedure for RET-zero and RET-span 

 

For  calibrating  the  retransmission  output,  both  retransmission  zero  and  retransmission 
span has to be calibrated. At a time there can be either one or two retransmission output 
available.  If  only  one  retransmission  is  used,  then  calibrate  RTZ1  (retransmission  1 
zero)  and  RTS1  (retransmission  1  span)and  if  there  are  two  retransmission  outputs 
available, then calibrate RTZ2 (retransmission 2 zero)and RTS2 (retransmission 2 span) 
for second retransmission. No need to feed input while calibrating retransmission o/p.it 
is  like  calibrating  using  digital  trim  pot.  Only  look  at  the  output,  display  value  has  no 
significance with output generated. 

 

Example:- 
At  retransmission  zero  calibration,  expected  output  is  4.00mA  and  it  gives  4.153mA. 
Then adjust the output value to 4.153 

by using ‘

IN

CREMENT KEY’. Now press ‘SET1’ to 

store  the  calibration  parameter  in  non-volatile  memory.  Similarly  one  can  calibrate 
retransmission span. 
 
 
 
 
 
 
 
 
 

 
 

Содержание LC5248E-AT

Страница 1: ...h A AU UT TO O T TU UN NE E L LC C5 52 29 96 6 A AT T L LC C5 52 24 48 8E E A AT T L LC C5 52 29 96 6V V A AT T Masibus Automation Instrumentation Pvt Ltd B 30 GIDC Electronics Estate Sector 25 Gandhi...

Страница 2: ...nmental Conditions 9 3 PHYSICAL SPECIFICATIONS MOUNTING DETAILS 9 3 1 LC5296 AT 9 3 3 LC5248E AT 9 3 4 LC5296V AT 10 4 TERMINAL CONNECTIONS 11 4 1 LC5296 AT 11 4 3 LC5248E AT 12 4 4 LC5296V AT 12 4 4...

Страница 3: ...N PROCEDURE 36 9 1 Procedure for CAL zero and CAL span 36 9 2 Procedure for RET zero and RET span 36 10 COMMUNICATION PROTOCOL MODBUS RTU 37 10 1 Introduction 37 10 2 Parameter Address Details 37 10 3...

Страница 4: ...his manual are the trademarks or registered trademarks of Masibus Automation and Instrumentation P Ltd herein after referred to as MASIBUS Adobe Acrobat and Postscript are either registered trademarks...

Страница 5: ...nter 5296 sub menu will come COP it has two options PID or ONOF 2 Selected COP in above menu will be displayed on power on condition Like PV 5296 SV TC E then PV COP SV ONOF If COP PID is selected the...

Страница 6: ...ng resistance for RTD Maximum 15 ohms wire Conductor resistance between three wires should be equal NMRR 40 dB CMRR 120 dB Temp co 150ppm C Input Impedance 1M Voltage Input 250 Current Input Max Volta...

Страница 7: ...Output 2 Relays Relay 1 For PID or ON OFF Controlling Used as Alarm 1 Output if Output Type is Linear Relay 2 Alarm 2 Output SSR Output Voltage Pulse Output Available at Terminals of Relay 1 Linear Ou...

Страница 8: ...DC 2A resistive load Loop Power Supply Supply Voltage 24VDC 1V 26mAwith Inbuilt Short Circuit Protection Minimum load resistance 800 ohms 2 4 Communication Details Communication Interface RS485 2 Wire...

Страница 9: ...erature 0 to 55 C Storage Temperature 0 to 80 C 3 PHYSICAL SPECIFICATIONS MOUNTING DETAILS 3 1 LC5296 AT Front Bezel 96 x 96 mm Panel Cutout 92mm 0 8 x 92mm 0 8 Depth Behind The Panel 65 mm withTermin...

Страница 10: ...of 44 Fig 3 3 Mounting Details for LC5248E AT 3 4 LC5296V AT Front Bezel 96 x 48 mm Panel Cutout 92mm 0 8 x 45mm 0 6 Depth Behind The Panel 75 mm with Terminal Weight 300 g Approx Enclosure Material M...

Страница 11: ...t slide wire feedback Terminal 3 4 SSR Pulse o p 5 NC2 6 NO2 7 C2 For Relay 2 potential free Contacts Use 230V 2A load Alarm 2 o p Reverse Relay if output type is Motor position control without slide...

Страница 12: ...Power Supply Input 8 C1 9 NO1 10 NC1 For Relay 1 potential free Contacts Use 230V 2A load On Off Control o p 11 LPS 24VDC Loop power supply Terminal 14 is ground Reference 12 D 13 D Modbus RTU Communi...

Страница 13: ...ons Wiring must be suitable for Voltage Current and temperature rating of the system Provide power from a single phase instrument power supply If there is a lot of noise in the power line insert an in...

Страница 14: ...Panel Description of LC5296 AT Name of Part Symbol Function Increment Key Increment the Value of any Parameter Shows ambient value for T C Input in RUN mode In Manual Mode this key is used to Incremen...

Страница 15: ...PV 4 digital 0 56 inch RED Display Display process value Display parameter name when user set parameter Display error message when an error occurs SV Set Value Display SV 4 digital 0 4 inch GREEN Dis...

Страница 16: ...n key is A M Shows remaining soak time when pressed if function key is selected SOK T PV Process Value Display PV 4 digital 0 40 inch RED Display for LC5248E AT 4 digital 0 36 inch RED Display for LC5...

Страница 17: ...Model LC5296 AT LC5248E AT LC5296V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 17 of 44 6 MENU LAYOUT 6 1 Menu Layout for LC5296 AT...

Страница 18: ...Model LC5296 AT LC5248E AT LC5296V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 18 of 44...

Страница 19: ...Model LC5296 AT LC5248E AT LC5296V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 19 of 44 6 2 Menu Layout for LC5248E AT LC5296V AT...

Страница 20: ...Model LC5296 AT LC5248E AT LC5296V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 20 of 44...

Страница 21: ...Shows only if Symbol Name C1 SP C1 sp Control Set Point 1 Range Depending on PV sensor type selected 100 A1 SP A1 sp Alarm Set Point 1 Range Depending on PV sensor type selected 100 Output Type is LIN...

Страница 22: ...sed to decrease effect of D term in PID output for some critical operating condition 0 01 to 1 00 1 00 Control Type COP is PID SMP smp Sampling Period Set Sampling Period 200 500 1 0 200 ms 1 500 ms 2...

Страница 23: ...t Value Input type range RTD Thermocouple 100 0 C to 100 0 C Linear 1000 to 1000 0 0 FLTR fltr Filter Enable or Disable Filter for PV Input yes no 0 NO 1 YES 1 YES TSP1 tsp1 Type of Set Point Set Type...

Страница 24: ...38400 bps 0 9600 RT 1 rt 1 Retransmission 1 Type Retransmission 1 Output Type This output is according to PV input Zero Span acts as Min Max value of retransmission o p scale respectively 0 5v 1 5v 0...

Страница 25: ...hen CO LO 100 0 Control output type is PID A1TP a1tp Alarm Type 1 Refer ALARM Type Table on Page 39 0 to 15 6 PV A L Output Type is LIN Linear A2TP a2tp Alarm Type 2 Refer ALARM Type Table on Page 39...

Страница 26: ...F CJC Fix cold junction Compensation Set Fix cold junction Compensation value 0 to 60 0 C 0 0 Input Type is TC A CJC is NO F KEY F KEY Function Key Select A M or Soak time A M SOK T 0 A M 1 SOK T SOK...

Страница 27: ...ission 1 output selection If Retransmission 1 type output is 4 20mA then the linear output will function respectively No need to feed input while calibrating Rx o p just make the value in calibration...

Страница 28: ...r operations On Off hysteresis prevents the output from chattering or making fast continual switches if the cycling above and below the set point occurs very rapidly Figure 7 1 Typical Relay operation...

Страница 29: ...action is the action which the control output varies in proportion to the deviation between the setting value and the processing temperature If the proportional band is narrowed even if the output cha...

Страница 30: ...n the proportioning band but likely not on the set point Some controls use manual reset as a digital user programmable value this allows the user to preprogram the approximate output requirement at th...

Страница 31: ...e sudden change of set point The ramp function is performed in following conditions The target set point is changed Target set point number is changed For example Switching from SP 1 to SP 2 The power...

Страница 32: ...ully closed position to the fully open position With the preset operating time the controller controls the valve by estimating the position There is no need for feedback input wiring When in manual op...

Страница 33: ...Ref figure 5 4 Pv d b Deviation High Low Band alarm Ref figure 6 5 Pv a H Absolute value High alarm Ref figure 1 6 Pv A L Absolute value Low alarm Ref figure 2 7 SP A H Absolute value set point high...

Страница 34: ...248E AT LC5296V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 34 of 44 NOTE LIT LED on UNLIT LED off Up arrow indicate Alarm will ON from this value Down arrow indicate Alarm will O...

Страница 35: ...behave exactly the opposite However it s worth mentioning that the relays will be in off de energized state on Power on reset condition They will energize only after approximate 5 seconds When alarm t...

Страница 36: ...are calibrated Here one shot calibration technique is used i e the zero and span are calibrated once Individual zero or span can also be calibrated first calibrate zero and then span While calibrating...

Страница 37: ...egister Use to write programmable register The error checking field contains a 16 bit value implemented as two eight bit bytes The error check value is the result of a Cyclical Redundancy Check CRC ca...

Страница 38: ...2 hysteresis 40025 Int R W 26 Alarm 1 logic 40026 Int R W 27 Alarm 2 logic 40027 Int R W 28 Alarm 1 delay 40028 Int R W 29 Alarm 2 delay 40029 Int R W 30 Auto CJC 40030 Int R W 31 Fixed CJC 40031 Int...

Страница 39: ...in the query is not an allowable address for the salve 03 Illegal data value A value contained in the query data field is not an allowable value for the salve 06 When Master device write some paramet...

Страница 40: ...44 11 APPENDIX 11 1 Troubleshooting If the operating display does not appear after turning on the controller s power follow the measures in the procedure below If a problem appears complicated contact...

Страница 41: ...E ACTION OPEN OVER UNDER 4 20mA PV UP DIR 20 8 20 8 3 2 PV DOWN REV 3 2 3 2 20 8 PV UP REV 20 8 3 2 20 8 PV DOWN DIR 3 2 20 8 3 2 1 5V PV UP DIR 5 2V 5 2V 0 8V PV DOWN REV 0 8V 0 8V 5 2V PV UP REV 5 2...

Страница 42: ...ailable There are Two Addon Card Slots available on PCB of Signal Card The Right Most Slot is fixed for Retransmission Analog Output And the other Left Slot can be used for either Retransmission or RS...

Страница 43: ...96V AT masibus Doc Ref no m61C om 301 Issue no 13 User s Manual Page 43 of 44 Jumper Setting for Retransmission card m61Cao102 Applicable for LC5248E AT Jumper Setting for Retransmission card m49Acp10...

Страница 44: ...44 of 44 11 6 Load connection Electrical precautions during use Electrical noise generated by switching of inductive loads can create momentary disruption erratic display and latch up data loss or pe...

Отзывы: