User Manual
P a g e
|
31
of
49
Auto Stepping
To Enable Auto Stepping, select Source Type as STEP & Un-
Check the Manual CheckBox.
If this mode is enabled,
(Step UP) or
(Step Down) icon will
appear in Source Display Window in RUN Page and F1 & F2
Button change to
START
&
SETTING
respectively.
Automated Step can be started by Pressing
F1
key (START). After
that F1 & F2 key will change to
PAUSE
&
STOP
respectively. So by
pressing F1 & F2 key running STEP can be PAUSE or STOP at
any time in RUN Page.
STEP Setting can be accessed directly by
F2
key (SETTING).
NOTE: While STEP is running
STEP settings can’t be accessible
and Source Page Parameter settings can’t be change.Stop STEP
before changing any settings.
4.2.2 RAMP Page
Parameter Name
Description / Options
Hold@0%(s)
Time to wait at Low(0%) level in second.
This parameter is use for
Repeat FormatUP/DOWN
or
DOWN/UP
.
Range:
0 to 9999
Rise Time (s)
Time to Increase from Low to High Level.
Range:
1 to 9999