User Manual
P a g e
|
43
of
49
Apply mA Input value near to Recommended Zero Value for Sub
Range 1 (for mA sub range 1 it is 1.000mA) from reliable source
device.
For example, If apply 1.000mA from the external source.
Measured
value shows the value that has been measured by the
LC-11. If this value is 1.0200 enter 1.0000 value in
ZERO Actual
Value
Edit Box & Press
MENU/ENT
key to calibrate the ZERO.
Similarly, for
SPAN for Sub Range #1
calibration,
Apply mA Input value near to Recommended Span Value for Sub
Range 1 (for mA sub range 1 it is 11.000mA) from reliable source
device.
For example, If apply 11.000mA from the external source. If the
Measured
value shows 10.9950 enter 11.0000 value in
SPAN
Actual Value
Edit Box & Press
MENU/ENT
key to calibrate the
SPAN.
Follow the same procedure to calibrate for the Sub Range 2.
The same procedure is applicable for other Input Type.
To calibrate input use other source in series or parallel to device
according to the input type which is ten times accurate compared
to the range of the instrument
.
Note
: Calibration of mA Input will also calibrate mA(24V) input. No
need to
separately calibrate mA(24V) Input.
Procedure for calibration of SourceMode
To enter into the Source Calibration, Select the
SOURCE
option and
press
F1
key to see Source calibration page. To calibrate the Source
mode
ZERO
and
SPAN
both has to be calibrated. First calibrate Zero
and then Span.
First select the Output Type which to be calibrated.