Example: - Calibrating mA Output
To calibrate
ZERO,
Select
ZERO Measured
Edit Box, when this Edit Box is selected
LC-11 will source value that is seen in
ZERO Expected
Value
(here LC-11 will source 4.000mA).
Now Measure the source value in Reliable Measure Unit. For
Example the external measure unit is measuring 3.9980mA. Then
enter 3.9980 value in
ZERO Measured
Edit Box & Press
MENU/ENT
key to calibrate the ZERO.
To calibrate
SPAN,
Select
SPAN Measured
Edit Box, when this Edit Box is selected
LC-11 will source value that is seen in
SPAN Expected
Value
(here LC-11 will source 20.000mA).
Now Measure the source value in Reliable Measure Unit. For
Example the external measure unit is measuring 20.0100mA.
Then enter 20.0100value in
SPAN Measured
Edit Box & Press
MENU/ENT
key to calibrate the SPAN.
Follow the same procedure to calibrate the other output types.
Expected Values of Zero and Span for source calibration are as per
below table.
Output Type
Expected Zero Value
Expected Span Value
mA
4.0000
20.0000
mV
25.000
240.000
V
1.0000
11.0000
Note
: Calibration of mA Output will also calibrate mA(2W) Output.
No need to
separately calibrate mA(2W) Output.