Model: MC-1-M (2U)
Doc. Ref. no. : m05/om/101
Issue no. 14
User’s Manual
Page 56 of 103
10.2.5 COM2 Serial Time Frame Command:
•
Command: TC2x [Serial frame Command]
Applicable Options: TC21 / TC22 (Please refer ‘H’ Command menu for applicable meanings of
mentioned options)
The COM2 of GPS Master Clock Model MC-1-M allows the time format in two different ways: NGTS
format and T-format. These formats are already explained before. These commands are:
TC21: NGTS mode
TC22: T-format mode
Remember that T-format sends the time frame each second while the NGTS sends the time frame each
minute.
10.2.6 Event Command:
•
Command: Ex [PPM / PPH Event Command]
Applicable Options: E1 /E2 (Please refer ‘H’ Command menu for applicable meanings of mentioned
options)
The pulse event signal is available at the terminal strip on the rear panel of GPS Master Clock Model MC-
1-M. This signal triggers relay either every minute or every hour based on the configuration set by the
command as listed below:
E1:
1 Minute mode
E2:
1 Hour Mode
10.2.7 Additional Event’s Command:
The additional four event signals are available at the terminal strip on the rear panel of GPS Master Clock
Model MC-1-M. User can configure these opto-relays through COM2 for time ranging from 1 to 86400
seconds individually for separate event outputs. User can configure their Time period as well as their ON
time through COM2 port.
Time Period limit:1 to 86400 seconds
(0 values are to inhibit/stop particular event output).
ON time limit:50 milliseconds (min.) to 50% (max.) of particular event time period. Ensure that ON time
value of all additional events are to be entered in milliseconds.
(If time period is 0, the ON time will be also 0 by default).
Whenever the new time period is set, the new event timing counter will start from the very next minute
and the contact will be energized after the settable time for that particular event. Please ensure to switch
power of the instrument OFF and ON once if the event time period settings are changed.
•
Command: ETx
Applicable Options: ET1 / ET2 / ET3 / ET4 (Please refer ‘H’ Command menu for applicable meanings
of mentioned options)
This command is used to configure Time Period of Additional Events. These Events can be configured to
trigger at every second to every 86400 seconds (24 hr.). These configuration commands are listed below: