GC-APPI Interface
Page 6
Installation of the GC-APPI Interface
Before installing the GC-APPI-Interface you have
to remove the ESI sprayer. Make sure the mass
spectrometer is set to “Off”, and the heated ion
transfer tube is set to room temperature (see
Figure 1 and 4).
Caution:
Removing the ESI sprayer while the
instrument is on or in standby will load the default
Tune file and eventually start the capillary heating.
Figure 1:
Instrument in “Off” mode
Disconnect the waste hose from the bottom of the
TNG source, as well as all gas, syringe, and
electrical connections.
Figure 2:
Instrument with ESI sprayer