GC-APPI Interface
Page 19
Figure 27:
Example of MS tuning in APPI mode
Setup the Instrument Method
The following is valid when using the Thermo
Scientific Trace 1310 GC. For any other GC, we
refer to its dedicated manual.
Figure 28:
Instrument Method Setup
Open the Instrument Method editor from the
Xcalibur Home page. In the editor, you will see the
different instruments.
Select the Trace 1310 GC and setup the desired GC
method. On the Aux Temperatures tab you have
to enter the temperature of the transfer line and
the temperature of the APPI source (listed as Aux.
Heater 2, see Figure 29). Both temperatures
Содержание MC513000