GC-APPI Interface
Page 20
should not differ a lot from each other. Typical
temperatures are between 225 and 300 deg. C.
Please note:
The maximum temperature of the APPI source is
330 deg. C.
For details of the GC setup and/or programming,
we refer to the GC Manual.
Figure 29:
Aux Temperature for Trace 1310 GC in
Instrument Method Setup
Switch to the MS to select or setup the
appropriate MS method with the corresponding
MS Tune file. If an autosampler is in use, also
setup the correct autosampler method. Save the
Instrument Method and setup a sequence file.
Setup the Sequence File
From the Xcalibur Home page, select the
Sequence editor to setup one or more sample
measurements. For details, we refer to the
standard MS operation manual.
Running a GC sample
When all instrument devices in Xcalibur are
“Ready for download”, you can start the sample or
sequence measurement from the sequence setup.