Promote your assets
5
Diagram 1-1 Detection Principle of SF
6
Gas Sensor
1.4
1.4
1.4
1.4
Application Place
I. The instrument is special products for detecting the SF
6
gas leak (without
Explosion-proof Certification), widely used in many occasions of using SF
6
gas
such as the electric power industry, metallurgy industry, petroleum chemical
industry etc.
II. Working environmental requirements: environmental temperature -20
℃
℃
~ + 60 ,
humidity 0 ~ 95% RH (no condensation)
Warning
Due to the mirror inside the measurement sensitive components, once the
mirror is defiled, the accuracy of measurement will be affected greatly,
so ban to use the product in dirty harsh environments.
1.5
1.5
1.5
1.5
Executed standards
GB12358-2006 General Technical Requirements for Gas Detection Alarms in
Workplace Environments
DL/T 846.6-2004 General technical specifications for high voltage test equipments
Part 6:SF6 gas leak detector
JJF 1263—2010 Calibration specification for the alarmed detector of sulfur
hexafluoride
Содержание MSF6-1000