Figure 3-25 Searching the sensor information Figure 3-25 Display the sensor address
View the current program version
As shown in Figure 3-27, select SOFT_VER, press the confirm key to entering
the interface of the software version information.
Figure 3-27 Select the SOFT_VER Figure 3-28 Software version information
3.3.3
3.3.3
3.3.3
3.3.3
Calibration Operation
The accuracy of the measured data is closely related to the calibration. Select the
“INST_CAL " in the main menu and press confirm key to entering the instrument
calibration submenu as shown in Figure 3-30. Set" CAL_LOCK " to open state before
instrument calibration as shown in Figure 3-31,and then to operate the instrument
calibration steps. It includes the single point calibration, two-point calibration (low
point and high point calibration) and zero point calibration.
The two-point calibration shall be operated after completing the zero point
calibration, turn into the air and the standard gas approximate the full scale to
calibrate the low point and high point.
The single point calibration is for improving the accuracy of the measuring value
at one point and the near points. It shall be done based on the accurate of two-point
calibration, so the single point calibration is meaningless without the accurate