Altitude
The product has been evaluated for operation up to a maximum of 6500 ft (2000 m), in accordance with
safety standards. Higher altitudes pose no safety risk, but instrument performance may be reduced.
NOTE
: For storage or shipping the allowable temperature range is -40 to 70°C and the allowable humidity
range is 5-95% non-condensing. After instrument exposure to extremes of temperature or
humidity, allow 2 hours for return to the recommended ranges before switching on.
1.6
Technical Specifications
1.6.1
Physical properties
Height:
40 cm (15.75”)
Width:
16 cm (6.3”)
Length:
51 cm (20”)
Mass:
16 kg (35.2 lbs)
1.6.2
Electrical properties
Maximum Power:
650 W
Line voltage:
100-240 V (automatically selected)
Frequency:
50/60 Hz
Input inrush current (A):
<40 A (cold start)
1.7
Technical support contact details
In the first instance please contact your distributor. If they are unable to resolve your query, please contact
Markes International (details below).