![Markes International Unity 2 Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/markes-international/unity-2/unity-2_installation-manual_1727983021.webp)
QUI-1056
Page 21
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
E:
www.markes.com
6.
Remove the trap alignment tool.
NEVER SWITCH UNITY 2 ON WITH THE TRAP ALIGNMENT TOOL INSTALLED
7.
Holding onto the glass collar of the cold trap, gently push the trap tube into the cold trap box.
DO NOT APPLY EXCESSIVE FORCE TO THE QUARTZ COLD TRAP TUBE
8.
As the cold trap contacts the o-ring seal at the heated valve end there will be an increased resistance
to movement. The trap must be pushed in a further 2-3mm to ensure it is correctly seated in this o-
ring.
NOTE
: Once the trap is properly inserted the glass collar should be approximately 1 mm from the
black plate
9.
Replace the Trap Pneumatic Assembly.