Series TT05 Digital Torque Testers
User’s Guide
20
To save the calibration information, select “Save & exit”. To exit without saving the data select
“Exit without saving”.
14. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
The calibration weight does not match the set value.
The entered calibration point is too close to the previous point.
CALIBRATION
COUNTER-CLOCKWISE
Load too close
to previous.
Please try again.
CALIBRATION
Load not stable.
Please try again.
CALIBRATION
Units must be lbFin.
Please try again
Press ENTER.
CALIBRATION
COMPRESSION
Load too low.
Please try again.
CALIBRATION
COMPLETE
Save & exit
Exit w/o saving