Series TT05 Digital Torque Testers
User’s Guide
9
Any combination of the above functions may be selected.
If tones are enabled, a tone will sound when the output, storage, and zero functions have occurred.
9.2 Break Settings
Select
Break Settings
from the
Break Detection
menu to configure the settings. The display appears as
follows:
9.3 Auto Output Settings
Scroll to
Auto Settings
in the
Break Detection
menu and press
ENTER
to configure the auto output
settings. Any combination may be selected. The display appears as follows:
Function Description
Enabled
Arms the break detection function. When enabled, the letter “B” appears on the
home screen, between the Mode and Unit indicators. Refer to the
Home Screen
and Controls
section for details.
Note:
Break Detection
and
First / Second Peak
mode cannot be enabled
simultaneously.
Break Settings
Refer to the following sub-sections for details.
Auto Output
Auto Storage
Automatically stores the peak reading to memory.
Auto Zero
Automatically zeroes the display following data transmission and/or storage. A
time delay may be configured in
Break Detection Settings
. Refer to the next
sub-section for details.
Threshold
Sets the percentage of full scale at which the break detection function becomes
active. This threshold is provided to ignore peaks that can occur during sample
loading and unloading.
Available settings: 1–90%, in 1% increments until 5%, 5% increments thereafter
.
% Drop
Sets the percentage drop from the peak reading at which the break is detected.
Available settings:
5%–90% in 5% increments.
Auto Zero Delay
Sets the time delay before the primary and peak readings are zeroed. Auto zero
can be disabled if required. Refer to the
Auto Output Settings
sub-section for
details. Available settings:
1–10 sec. in 1 sec. increments, and 10–60 sec. in 5 sec.
increments.
BREAK DETECTION
SETTINGS
Threshold: 5 %
% Drop: 50 %
Auto Zero Delay
5 sec.
BREAK DETECTION
Enabled
+ Break Settings
+ Auto Output
Auto Storage
Auto Zero