Series TT05 Digital Torque Testers
User’s Guide
10
Parameter Description
RS232/USB Output
Automatically output the peak when the break (% Drop) is detected.
Mitutoyo Output
Automatically output the peak when the break (% Drop) is detected.
Output Pin
Automatically toggle the SP1, SP2, or SP3 pins (active low). If not required, select
“NONE”.
10 FIRST / SECOND PEAK
This function captures two consecutive peaks which can occur over the course of a test, such as with
tamper-evident breakaway band closures, torque tools, and other samples.
10.1 Configuration
Several functions can be triggered automatically upon first and second peaks:
1. Transmit the first peak reading
2. Transmit the second peak reading
3. Save the first peak value to memory
4. Save the second peak value to memory
5. Zero the primary and peak readings
These automatic functions can help automate and expedite testing processes. If tones are enabled, an
audible tone will sound when the output, storage, and zero functions have occurred. In order for First /
Second Peak detection to be active, the appropriate operating mode must be selected. See the
Operating Modes
section for details. The display appears as follows:
Any combination of the above functions may be selected.
Function Description
Enabled
If enabled,
2PK
appears as one of the operating modes. In the main display, the
Peak readings will reference the first and second peaks – first peak on top,
second peak below. Refer to the
Home Screen & Control
section for details.
Peak Settings
Press
ENTER
to access the
Peak Settings
sub-menu. See the following sections
for details.
Auto Output
Press
ENTER
to access the
Auto Output Settings
sub-menu. See the following
sections for details.
Auto Store PK1
Automatically stores the first peak reading to memory.
Auto Store PK2
Automatically stores the second peak reading to memory.
Auto Zero
Automatically zeroes the display following data transmission and/or storage.
AUTO OUTPUT
SETTINGS
RS232/USB Output
Mitutoyo Output
Output Pin: NONE
FIRST/SECOND PEAK
*
Enabled
+ Peak Settings
+ Auto Output
* Auto Store PK1
Auto Store PK2
* Auto Zero