242
Servo
…
User´s manual
US edition
Order No: 66 00 261
11
Infant
Adult
Universal
Options
The system
General
The device complies with requirements of Medical
Device Directive 93/42/EEC.
Standards
EN IEC 60 601-1 (Class 1, Type B).
IEC 60601-2-12
EN 794-1
Electromagnetic compatibility (EMC)
According IEC 60601-1-2, 2nd edition (2001)
Immunity: Extended test to 30V/m
The EMC declaration, Information to the responsible
organization is available from MAQUET.
Patient range Adult (weight: 10-250kg)
Infant (weight: 0.5-30kg)
NIV (PC+PS) Infant range 3-30kg.
NIV Nasal CPAP Infant range 0.5-10kg
Operating conditions
Operating temperature range +10 to +40 ºC.
Relative humidity 15 to 95% non-condensing.
Atmospheric pressure 660 to 1060hPa.
Lowest pressure in patient circuit - 400 cmH
2
O.
Impact
Peak acceleration: 15 g.
Pulse duration:
6 ms.
Number of impacts:1000.
Non-operating conditions
Storage temperature -25 to +60 ºC (-13 to 140 ºF).
Storage relative humidity < 95%.
Storage atmospheric pressure 470 — 1060hPa.
Power supply
Power supply, automatic range selection
100-120V ±10%, 220-240 V ±10%, AC 50-60Hz.
Battery backup
2– 6 battery modules rechargable 12 V, 3.5 Ah each.
Recharge time approximately 3 h/battery. Battery
backup time approximately 3 h, when using 6 batteries.
External 12V DC 12.0V - 15.0V DC, 10A
Caution: : When ex12 V DC is used, at least one
installed Battery module is required to ensure proper
operation.
Max power consumption
At 110-120V: 2A, 190VA, 140W.
At 220- 240V: 1A, 190VA, 140W.
The ventilator
General
Dimensions User Interface: W 355 x D 53 x H 295 mm,
Patient Unit: W 300 x D 205 x H 415 mm.
Weight Approximately 20 kg
(User interface 5 kg, Patient Unit 15 kg).
Method of triggering Flow and pressure.
Gas supply
The gases supplied must be free from water, oil and
particles and other contaminants.
Air:
H
2
O < 7 g/m
3
Oil < 0.5 mg/m
3
Chlorine: Must not be detectable
1
Oxygen:
H
2
O < 20 mg/m
3
Inlet gas pressure 2 – 6.5 kPa x 100 (29 – 94 PSI).
Connection standards available AGA, DISS, NIST or
French.
Patient system gas connectors
Conical fittings Male 22 mm and female 15 mm.
In accordance with ISO 5356-1.
Gas exhaust port Male 30 mm cone.
User Interface
Weight approximately 5 kg. Can be attached to the
mobile cart, a table, railing or pipe (15 - 30 mm
diameter).
1.
If the compressed air is generated by a
liquid ring compressor there is a potential
risk of chlorine in the supplied air.
Technical data
Содержание SERVO-i V3.1
Страница 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Страница 2: ......
Страница 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Страница 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Страница 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Страница 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Страница 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Страница 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Страница 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Страница 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Страница 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Страница 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Страница 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Страница 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Страница 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Страница 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Страница 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Страница 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Страница 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Страница 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Страница 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Страница 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Страница 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Страница 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Страница 279: ......
Страница 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...