• To avoid fire hazard, keep all sources of
ignition away from the Servo-i Ventilator
System and oxygen hoses. Do not use
oxygen hoses that are worn, frayed, or
contaminated by combustible materials
such as grease or oils. Textiles, oils, and
other combustibles are easily ignited and
burn with great intensity in air enriched
with oxygen. Immediately disconnect the
ventilator from the oxygen supply, facility
power, and backup sources if there is a
smell of burning.
General cautions
• As a general rule always avoid contact
with external electrical connector pins. It is
recommended to have the module
compartment filled up with empty modules
to protect from spillage and dust.
• Federal law in the USA restricts this device
to sale by or on the order of a physician (or
a properly licensed practitioner).
• The Servo-i Ventilator System must be
serviced at regular intervals by specially
trained personnel. The intervals are stated
in the chapter Regular maintenance. Any
maintenance must be noted in a log book
for that purpose in accordance with
national regulations.
• MAQUET has no responsibility for the safe
operation of the equipment if service or
repair is done by a non-professional or by
persons who are not employed by or
authorized by MAQUET. We recommend
that service is done as part of a service
contract with MAQUET.
• MAQUET has no responsibility for the safe
operation of the equipment if the
equipment is used for anything other than
its intended use, as specified in this User´s
manual.
• A resuscitator should always be readily
accessible for extra safety.
• When connected to a patient, the system
must never be left unattended.
• The nebulizer module is inoperative when
the ventilator is running on batteries, to
reduce the power consumption.
• The Expiratory cassette must not be lifted
up when the ventilator is in operation. This
may, however, be done when in Standby
mode.
• Always use heat and moisture exchanger
(HME) or equipment to prevent
dehydration of lung tissue.
• Refer to the Installation instructions to
assemble the system or options to obtain
a proper mechanical assembly.
• When lifting or moving the ventilator
system or parts of the system, follow
established ergonomic guidelines, ask for
assistance, and take appropriate safety
precautions.
• Antistatic or electrically conductive
breathing tubing should not be used with
this lung ventilator system.
• Any scavenging system (Gas evac)
connected must comply to ISO8835-3
with regard to subatmospheric pressure
and induced flow. Otherwise ventilator
functions and patient safety may be
degraded.
• It is not recommended to use the Servo
Evac 180 in the Nasal CPAP mode.
• Values measured at the signal outputs of
the Servo-i Ventilator System and which
have been processed in auxiliary
equipment must not be used as a
substitute for therapeutic or diagnostic
decisions. Such decisions can be made
only by staff with medical expertise,
according to established and accepted
practice. If auxiliary equipment that has
not been delivered by MAQUET with the
system is used, MAQUET denies all
responsibility for the accuracy of signal
processing.
Warnings, cautions and important
Содержание SERVO-i V3.1
Страница 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Страница 2: ......
Страница 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Страница 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Страница 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Страница 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Страница 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Страница 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Страница 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Страница 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Страница 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Страница 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Страница 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Страница 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Страница 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Страница 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Страница 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Страница 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Страница 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Страница 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Страница 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Страница 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Страница 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Страница 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Страница 279: ......
Страница 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...