STARTING
W A R N I N G !
Before starting up check whether the mains power supply voltage
corresponds with the information stated on the nameplate.
W A R N I N G !
Start up only the built-in heated probe tube because otherwise
there is a risk of burning!
W A R N I N G !
Check whether the housing of the heating controller and the
connection box of the heated sample tube are earthed. Earth
resistance < 0.3 Ohm.
•
Switch on mains power supply.
•
Then inner temperature of the sample tube has to be measured during the heat-up phase.
The maximum temperature reached is the value for the temperature class. This has to be taken
from the table below. Adjust desired temperature via controller.
Function and correctness of the temperature control (temperature
class) is in the sole responsibility of the installer and operator.
•
The nameplate contains no information about the temperature class because this is determined
by the temperature controller only.
Temperature class
Operating/sampling temperature
°C
Limiter
°C
T2
280 [536 °F]
290 [554 °F]
T3
185 [365 °F]
190 [374 °F]
T4
120 [248 °F]
125 [257 °F]
T5
85 [185 °F]
90 [194 °F]