12
SP30-H../EX2 | 1.02.00
www.mc-techgroup.com
Figure 3
Mounting of the heated sample tube SP30-H../EX2
13
ELECTRICAL CONNECTIONS
W A R N I N G !
When connecting the equipment, please ensure that the supply
voltage is identical with the information provided on the model type
plate.
N O T E !
Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal voltages
of up to 1000 V, together with the associated standards and
stipulations.
In any case we recommend the use of temperature resistant cable!
A main switch and matching fuse must be provided externally!
The main circuit must be equipped with a fuse corresponding to the
nominal current (over current protection); for electrical details see
technical data.
It is recommended to use the low temperature alarm. In case of an
alarm the flow can be stopped and the components downstream the
probe are safe for demage.
Function and correctness of the temperature control (temperature
class) is in the sole responsibility of the installer and operator.