Make sure, that the installation location is accessible for maintenance and service.
The IP protection class of the pump motor is indicated on the type plate.
Warning
Condensate accumulated inside the pump head destroys the pump.
Condensate can form, for example, if the pressure dew point is ex-
ceeded or if there is a malfunction in the sampling process. If conden-
sate is to be expected, then the following measures must be taken:
Turn the pump head downwards so that the gas connections point
downwards.
Mount the pump at the highest point in the system. If this is not
possible, route the outlet tubing away from the pump at an incline.
These measures allow condensate formed in the pump outlet to flow
out of the pump head and be safely drained.
Protect the pump from dust.
Protect the pump from vibrations and jolts.
Warning
Personal injury and/or damage to property because of vibration
In conjunction with adjacent components, vibration of the pump may re-
sult in crushing and/or damage to these components.
Make sure that vibrations of the pump do not result in hazards as-
sociated with adjacent components.
Protect the pump against contact and intrusion of foreign matter.