8
LA | 1.01.01
www.mc-techgroup.com
6
INTRODUCTION
The M&C liquid alarm units series
LA
®
are designed for stationary continuous operation and when
installed properly guarantee a long service life and a minimum of maintenance work.
7
APPLICATION
Liquid alarm sensors
LA1S
and
LA25S
are used in gas sample conditioning systems for monitoring gas
cooling and condensate drain-off devices in order to provide protection for downstream analysis
instruments. This straightforward monitoring device reliably signals the condensate
ingress in the event
of cooling or condensate drain-off equipment being defective and thus avoids expensive down times as
well as high repair costs for analysis instruments.
The electronic controllers
LA1.1
and
LA1.4
are used for feeding and signal processing of the liquid
alarm sensors
LA1S
and
LA25S.
In the event of alarm, the power for the sample gas pump or a solenoid valve is to be switched off in the
gas sample conditioning system.
8
TECHNICAL DATA
Sensor Series LA
®
LA1S
LA25S
Flow chamber
LS
LS25
Part No.
03E1001
03E3100
03E1111
03E1120
Connection cable
length
(ø 4.2 mm, 2-core)
4 m [13.1 ft]
4 m [13.1 ft]
Each additional meter of connection cable = Part No.
03E9001
, total max. 100
meters [328 ft]
Pressure
0 to 4 bar g****
0 to 25 bar g (up to 100 bar g on
request*)
Flow rate
0 to 1000 l/h
Operating
temperature
+80 °C *** [176 °F]
Gas connections
ø 12 mm
G 1/4" i
1/4" NPT a
ø 8 mm **
Stagnant space
Approx. 3 ml
Approx. 3 ml
Material of sample
contacting parts ***
Glass, platinum
PVDF, FPM
SS316Ti, PVDF
SS316Ti
Type of mounting
Clamping
attachment
Wall mount
Screw on
Tube
Mounting position
Vertical
Electrical conductivity
> 50 µS/cm
Weight
50 g
[approx. 0.11 lb]
60 g
[approx. 0.13 lb]
140 g
[approx. 0.31 lb]
100 g
[approx. 0.22 lb]
* other pressure ranges on request
** other dimensions on request
*** other materials or temperatures on request
**** max. 2 bar when mounted in universal filter
A Electronic controllers type LA-1.1 LA-1.4
Содержание LA Series
Страница 26: ...26 LA 1 01 01 www mc techgroup com Figure 8 Flow diagram LA1 1 115 230 V 50 60 Hz ...
Страница 27: ...www mc techgroup com LA 1 01 01 27 Figure 9 Flow diagram LA1 1 24V AC DC ...
Страница 28: ...28 LA 1 01 01 www mc techgroup com Figure 10 Flow diagram LA1 4 115 230V 50 60Hz ...
Страница 29: ...www mc techgroup com LA 1 01 01 29 Figure 11 Flow diagram LA1 4 24V AC DC ...