INSTALLATION INSTRUCTIONS AND DIMENSIONS
The liquid alarm sensor
LA...S
is used for detecting liquids and is to be installed behind the gas
dehydration devices, e.g. behind a gas cooler, but always on the deepest point of the installation. This
is provided by the universal filters
F
…-...-D
and the flow chambers
LS
or
LS25
.
The filters
F...-...D
and the flow chambers
LS
are designed for wall mounting and are mounted by using
2 fastening screws.
The operating position is exclusively vertical, only then an immediate alarm is guaranteed in case of
liquid ingress.
The outdoor installation of the valves can only be carried out without special protective measures, when
the ambient temperature in winter does not fall below the dew point temperature of the sample gas.
The gas inlet and outlet of the filter can be positioned right or left by moving the blue mounting bracket.
There is also a vertical inlet connection with a dummy plug. The flow chamber has the gas inlet vertically
and the gas outlet horizontally right or left, depending on the type of mounting.
N O T E !
During mounting, the gas inlet and gas outlet of the universal filter and
the flow chamber can be rotated 180 degrees.
The flow direction is marked with arrows.
The standard gas connections of the universal filters
F...-...-D
and the flow chamber
LS
are G 1/4" i.
Depending on the application, the gas connections can be made with soft tubing, e.g. PVC, Novoprene,
FPM or with rigid plastic or metal tubes, e.g. PTFE or stainless steel.
You will find the corresponding connection fittings in our product range.
Содержание LA Series
Страница 26: ...26 LA 1 01 01 www mc techgroup com Figure 8 Flow diagram LA1 1 115 230 V 50 60 Hz ...
Страница 27: ...www mc techgroup com LA 1 01 01 27 Figure 9 Flow diagram LA1 1 24V AC DC ...
Страница 28: ...28 LA 1 01 01 www mc techgroup com Figure 10 Flow diagram LA1 4 115 230V 50 60Hz ...
Страница 29: ...www mc techgroup com LA 1 01 01 29 Figure 11 Flow diagram LA1 4 24V AC DC ...