3.2
Urinal Bottle / Bedpan
–
Operating Cycle (Refer to laminated quick reference instructions)
1. Open door, using or press
2.
Set rack position and load items to be processed as shown in 3.1
3. Select URINE / BEDPAN cycle to close door & start cycle or
4. Close door, using either the
or press
5. When the door closes, to start cycle use the
or press
within 10 Seconds whilst
“START” is
flashing
Cycles stages will be shown on LCD display
1.
Flush
2.
Cold water wash
3.
Hot water wash with detergent
4.
Hot water rinse
5.
Disinfection
6.
Cool down
7.
Completed
8.
When the cycle is completed the display will flash
COMPLETED OPEN DOOR
Open door, use the
or press
NOTE
: If temperature inside the chamber is too high the display will flash,
WAITING COOLING
When the temperature has dropped to a safe level the display will flash,
OPEN DOOR
Open door, use
or press
To run the same cycle repeat steps 1 to 3
If you don’t
need to run another cycle, close door by pressing
NOTE:
•
In an emergency, pressing the Power button on the top left hand corner of the display will power off the device.
•
On powering up after a cancelled cycle the device will display a Fault code 904
–
Power Interrupted (see note below)
NOTE:
•
On powering up device will prompt to complete a
“Purge Reset”, if the condition that caused the fault has been
rectified press “Select” to start the “Purge Reset”. On completion of the “Purge Reset”
device will restart previous cycle