Bedpan / Urinal Bottle and Utensil / Bowl Washer Disinfector (WDS)
Operation, Maintenance and Installation Manual
Issue 16
Page
13
August 22
•
Mains ISO switch must be suitably placed so device will not obstruct safe access to the isolator once installed.
•
Ensure isolator is suitably placed so device will not obstruct safe access to the isolator once installed.
•
Device must be earthed.
•
Malmet recommends having a 30mA RCD in the mains supply fixed wiring.
•
If the supply cord is damaged, it shall be replaced by the manufacturer or its service agent or similarly qualified person in
order to avoid any potential hazard.
2.5
Commissioning
a)
Before switching on the device make sure the
DEVICE IS LEVEL
and
WATER TAPS ARE ON
. Check that the
DRAIN
WASTE
is connected.
b)
Turn on the power at the isolation switch and press the standby button on the front display to power on. The LCD
display will illuminate and go to standby mode.
c)
Check that the water tank is filling with cold water and that it fills to Level 3. Ensure the lid is put back on water tank.
d)
Check that the steam generator tank has filled to the full Level 3. The steam generator will preheat to 85°C in standby
mode or during cycle.
e)
Check that the 5 litre detergent bottle has been fitted. From the menu, run the DETERG/START to purge detergent
through the line to flow sensor. This may need to be done twice if the line is empty on start-up (see 5.10).
f)
Flush approximately 1 litre of water down the steam generator tank overflow pipe.
This will fill the ‘S’ Trap at the hose
junction and prevent steam coming back up into the water tank.
g)
From the Menu, run Purge Reset three times (see 5.9). Check for any water leaks.
h)
Open the door and turn power off at the control. Check racks for freedom of movement. Check that flush nozzles
rotate freely. Check that all spray nozzles are tight. Turn power back on and close door by pressing the door button.
i)
Select the BEDPAN/URINE cycle and let it run through the cycle. Repeat the same procedure for the URINE and BOWL
cycles.
j)
Replace all covers.
Note:
DO NOT USE THE DEVICE WITHOUT THE WATER SUPPLY TURNED ON