DEVICE SPECIFICATIONS
Operation
environment
Temperature
Relative Humidity
Ambient Pressure
Boot-up Time
Warm-up Time
+5 °C to +40 °C / + 41 °F to +104 °F
15 % to 93 % (non-condensating)
98 kPa to 104 kPa
depends on connected PC
< 1 min
Transport
&
Storage
environment
Storage Temperature:
Transport Temperature:
0
C to 50
C, 32 °F to 122°F
-25
C to 70
C, -13 °F to 158 °F
Storage and Transport
rel. Humidity
10 % to 95 % (non-condensating)
Altitude rating
Max. operating altitude 2000 m / 6561 ft. above sea level
Dimensions
120 mm x 93 mm x 30 mm /
4,7 in x 3,7 in x 1,2 in
Weight
142 g / 0,3 lb
Language
Settings
English,
English
(Australian),
German, Spanish, French, Italian,
Dutch (Belgium), Turkish, Chinese,
Japanese, Hungarian (Magyar), Russian
Data Interfaces
PC connection
USB
Connectors
USB
USB-B to PC
Power Supply
USB
Approximately 310 mA
ABR
Stimulus
Type
CE-CHIRP
®
(135 Hz to 7.5 kHz)
Level
35 dB nHL
Stimulus rate
88 /s left ear, 92.5 /s right ear
Transducers
IP30 ABR for EarCup
TM
IP30 ABR for ear tips
BERAphone
®
Test modes
Binaural (only with IP30) or monaural
Recording
Analysis time
Maximum 180 s of artifact-free data samples
A/D resolution
16 bit
Artifact reject
system
Peak rejection level
Artifact rejection
level
100 uV
Display
Test value diagram with line towards
PASS, signal quality or real time EEG
Measurement
frequency
183 Hz, 73 Hz