23/10/2019
SYMPTOM
POSSIBLE
CAUSE
SUGGESTED TROUBLESHOOTING
BERAphone
®
does not
make contact
well on the
baby's head
Different head
sizes
Depending on the size of the head the position of
the vertex electrode can be adapted by rotating
the black disc in which the electrode is mounted.
Moving of the
vertex electrode
does not allow
ideal contact
For this special case an extra long electrode is
available to replace the vertex electode. This can
be
ordered
(stainless-steel
electrode
for
premature babies).
Only for MB 11 Classic
Impedance
traffic light
remains red
Poor fit or
insufficient
skin
preperation
1. Check the contact of the electrode on the skin.
2. Check that the electrode lead wire is properly and
securely connected to the electrode.
3. Remove the electrode and prepare the skin again
with a skin preparation product such as Nuprep
®
or an electrode skin prep pad at the site, wipe it off
and then re-apply the electrode to the site.
High artifact
(poor
signal
quality)
during
the
screening
1. Check the state of the baby. A quiet, sleeping baby
is most conducive to a high quality screening. Calm
the baby with a pacifier if appropriate.
2. Position the baby so he/she is comfortable and
in a way to release any tension in the neck. A
rolle
d up blanket tucked to support the baby’s
back and in the curve of the back and neck can
help to release tension in this area.
3. Check that the electrodes remain in contact with
the skin. If any electrode has become detached
or loosened from contact, press it back into
place or replace it.
4. Click on the
EEG
checkbox on the measurement
window and view the EEG at the top of the screen.
If there is high amplitude noise or cyclical noise
evident in the EEG then there may be
electromagnetic
interference
in
the
test
environment. Try moving the system to a different
location in the room or into another room.
Screening
result is not
progressing
1. Check that the insert earphone eartip remains
securely inserted into the baby’s ear canal(s). If
you are using the EarCups
TM
, check that they
are in contact with the baby’s skin around the
entire circumference of the baby’s ear.
2. Check the state of the baby and quiet the baby
if possible.