Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
11
23/10/2019
To use the device, establishing a PC connection via USB is
required. See section 4.4.1 on how to safely establish a
connection with a power supplied PC or laptop (medical
device/non-medical device) or to a battery-driven laptop.
This equipment is intended to be connected to other equipment
thus forming a Medical Electrical System. External equipment
intended for connection to signal input, signal output or other
connectors shall comply with the relevant product standard e.g.
IEC 60950-1 for IT equipment and the IEC 60601-series for
medical electrical equipment. In addition, all such combinations
– Medical Electrical Systems – shall comply with the safety
requirements stated the general standard IEC 60601-1, edition
3, clause 16. Any equipment not complying with the leakage
current requirements in IEC 60601-1 shall be kept outside the
patient environment i.e. at least 1.5 m from the patient support
or shall be supplied via a separation transformer to reduce the
leakage currents. Any person who connects external equipment
to signal input, signal output or other connectors has formed a
Medical Electrical System and is therefore responsible for the
system to comply with the requirements. If in doubt, contact
qualified medical technician or your local representative. If the
device is connected to a PC (IT equipment forming a system)
ensure not to touch the patient while operating the PC.
The MB 11 must only be used with a medical PC or a battery-
driven laptop. If the device is connected to a PC (IT equipment
forming a system) assembly and modifications shall be evaluated
by qualified medical technician according to safety regulations in
IEC 60601.
The consequence of not following this warning could be a too
high leakage current to the patient.
When connected to computer equipment do not simultaneously
touch the computer equipment and the patient at the same time.
The consequence could be a too high leakage current to the patient.
The operator of the MB 11 BERAphone
®
/Classic must not
touch the electrode part of the MB 11 BERAphone
®
/Classic
during a measurement.
WARNING
WARNING
WARNING