Cleaning the Case and Cables
Use caution while cleaning.
Before cleaning, unplug MB 11 from USB connector.
Use a damp cloth to clean the plastic parts of the MB 11
BERAphone
®
/Classic.
If disinfection is required, use a disinfectant wipe rather
than a spray product. Make sure that excess liquid from
the wipe does not seep into any sensitive areas such as
connectors and seams where plastic pieces connect.
Follow the instructions on the disinfection product.
Cleaning and Disinfecting the BERAphone
®
3.3.3.1 General
The BERAphone
®
features stainless-steel electrodes and an ear cushion designed to
make direct contact with the baby’s skin. Assuming intact skin on the baby, the device
is considered non-critical according to the CDC guidelines for minimizing cross-
infection, meaning that the device contacts the patient only externally on intact skin.
Therefore, sterilization is not required. However, the device must be cleaned and
disinfected with a medical or hospital-grade disinfectant prior to re-use.
For a detailed description of cleaning and disinfection of the BERAphone
®
, see the
following sections 3.3.3.2 to 3.3.3.4.
3.3.3.2 Disinfectants
Use of a non-alcohol-based disinfectant is recommended. Non-alcohol-based products
contain the active ingredient referred to as quaternary ammonia compound. The
quaternary ammonia compound is specifically designed to disinfect rubber, plastic,
silicone and acrylic products which are commonly used in hearing evaluation devices.
Many common disinfectant wipes present in hospitals contain alcohol as a main
disinfection ingredient. However, alcohol chemically denatures certain materials, such
as the material used in the BERAphone
®
ear cushion. With repeated exposure to
alcohol-based disinfectants, the material of the cushion will harden, crack and
breakdown over time. The higher the alcohol content of the disinfectant, the faster the
ear cushion will be affected.
If alcohol disinfectant wipes are used to disinfect the BERAphone
®
, the earphone
cushion will need to be replaced more frequently than if a non-alcohol-based
disinfection product is used.
CAUTION