Cleaning and Disinfection Recommendations
It is recommended that parts (device and accessories like headphones, ear
cushions) which come in direct contact with the patient be subjected to
standard cleaning and disinfecting procedure between patients. Please
disinfect the instrument and its accessories in dependence on the specific
existing infection potential.
Recommendations for cleaning and disinfection of MAICO device presented in
this document are not intended to replace or contradict policies in effect or
procedures required for infection control at the facility.
If there is not a high infection potential, MAICO recommends:
Before cleaning always switch off and disconnect the device from power
supply.
For cleaning use a lightly dampened cloth with soap water solution
Disinfect the instrument and its accessories by wiping the surfaces with wet
Sani-Cloth
®
Active wipes or a comparable product and allow them to take
effect for the duration related to the specific disinfection aim listed in product
data sheet of the disinfection product. Please follow also its instructions for
cleaning.
Wipe before and after each patient
After contamination
After infectious patients
Disinfect computer, keyboard, transport trolley etc. with Sani-Cloth
®
Active
wipes:
once a week
after contamination
when polluted
To avoid damage of the device and its accessories, please
mind the following:
Do not autoclave or sterilize.
Do not use the device in the presence of fluid that
can come into contact with any of the electronic
components or wiring.
Should the user suspect fluids have contacted the system
components or accessories, the unit should not be used
until deemed safe by a MAICO certified service
technician.