Operation Manual MA 42
8100460 Rev.7 Ver.1
7
12/15
use of electrical medical devices in medical electrical
systems.
Any person who connects external equipment has
created a system and is therefore responsible for the
system complying with the requirements of EN 60601-
1-1. If in doubt, contact your service technician or local
representative for help.
The instrument is not intended for operation in areas
with an explosion hazard.
The terminals must under no circumstances be short-
circuited.
To avoid the risk of electric shock, this equipment must
only be connected to the medical power supply
originally delivered by MAICO. Using another power
supply can also lead to electrical damage on the
instrument.
In case of emergency, disconnect the instrument from
the supply mains by pulling the plug out of the power
socket.
Do not position the instrument in a way that it is
difficult to operate the disconnection device. The
supply mains and the power socket shall be accessible
at all times.
Do not use the instrument if the mains cable and/or
the outlet is damaged.
Prevent cable breakage: cables must not be bend or
buckled.
In order to maintain a high level of safety and to ensure
the instrument works properly, it is necessary to have
the instrument and its power supply checked
according to the medical electrical safety standard
EN 60601-1 by a qualified service technician at least
once a year. For more information please see Chapter
CAUTION
CAUTION
WARNING
WARNING
WARNING
Содержание MA 42
Страница 1: ...Operation Manual MA 42...
Страница 2: ......