Speech Calibration
To calibrate the MIC or CD speech input, select the input with the Signal
selector key (11) or (15). Press the Monitor button (6), the monitor setting
screen appears, press now the Incal button (19) on the right side of the
display and the calibration screen appears.
Play the reference signal with CD or talk into the microphone. Use the left or
right level control (1) and adjust the levels until the VU-meter show all yellow
and one green light. If one or more red lights are on, reduce the level using
the corresponding dials (1).
Store the calibration and leave the calibration mode with the Ok (6) button on
the left side of the display.
7.2.4
Documentation of the Results
Printout
Enter the User Menu by pressing the functional button (19) and press the
button Print (13) to printout the results.
Store Results as PDF
Enter the User Menu by pressing the functional button (19) and press the
button PDF (11) to save the results as PDF file on the SD memory card or USB
flash drive.
Note: SD memory card or USB must be inserted into the device for saving to
occur. System configuration designates where the results will be store and is
set to USB from the factory. It is recommended to store PDF results to USB for
easy transfer to a computer.
Store the Result in the Patient List
Enter the User Menu by pressing the functional button (19) and press the
button Patient (15) to enter the patient list. Enter a new patient by pressing
the button New Patient (15) or select a current patient by rotating the dial (1)
to highlight the patient name and press (2) to select. It is also possible to print
a result from the patient list or to create a PDF of the result. Select a stored
session of a patient in the patient list and press the button PDF (11) or Print
(13). In this case, the patient information will be included in the PDF or
printout. Press the New button (7) in the patient list, to start a new session.
New Session
To clear and create a new session, select the New button (7) from the tone
audiometry mode.