Operation Manual MA 33
8100092 Rev. 9
79
19/08/2019
6.2.2 Calibration Values and Maximum Levels
– Bone Conduction (MA 33 BC)
Calibration values: Bone conductor Radioear B71W
Force: 4.9 ... 5.9 N Mastoid placement
Coupler IEC 60318-6, ANSI 3.6-2010 and ISO 389-3
Frequency [Hz]
Reference
equivalent
threshold force
level for tone
Air radiation*
Max level
[dB] (re 1µN)
Min - max
[dB]
Tone
[dB HL]
vcvcvc
250
67.0
-
35
500
58.0
-
60
750
48.5
-
65
1000
42.5
-
70
1500
36.5
-
70
2000
31.0
-
70
3000
30.0
4-18
70
4000
35.5
-
70
6000
40.0
10.5-31
45