33
Magtrol Model DSP7000 Dynamometer Controller
Chapter 3 – Installation/Configuration
SETUP
5. Press SCALE I to add a filter if desired.
6. Press ShIFT. The display should appear as follows:
POWER
BRAKE STATUS
SET POINT
SET POINT
P
I
D
TORQUE
SPEED
USER DISPLAY
TSC1
MAX TORQUE
XXXX
Figure 3–26 TSC1 Setup Menu
7. Press the ToRqUE UNITS button and use
and
buttons and Decrease/Increase dial to
set maximum torque.
8. Press ShIFT 3 times to complete the setup and return to the main menu.
3.3
torque filter setup
Note:
Filter setup should take place after hardware installation and software
configuration of the chosen testing instruments has been completed.
See Chapter 3 – Installation/Configuration
.
There are six different Filter settings to choose from including 2 hz, 5 hz, 10 hz, 20 hz, 50 hz, 100
hz and NoNE. The following instructions show how to select the desired Filter for each channel:
1. Starting from the main menu, press ShIFT.
2. Press SETUP button.
3. Select DYNo.
4. Select DYNAMoMETER.
5. Press SCALE I button until the desired filter setting is reached.
6. Press ShIFT repeatedly until the main menu is reached. The number of times will vary
depending on testing instrument selection.
7. To setup a filter for TSC2, press TSC to switch to TSC2 setup and follow the instructions
above.
3.4
instrumentation setup (speed)
Note:
Regardless of the input selected for the speed, the speed calculated
is made available to the USb/gPIb/RS-232 control loop at a rate
of 488 samples per second and to the display at a slower rate of
about 1/4 second.
Tach A is designed to be used with a single output speed transducer. The maximum frequency into
the system when using the Tach A selection is 200 khz. The pulse per revolution (ppr) can be set
from 1 to 99999 ppr.