MAGSTIM BiStim2 Скачать руководство пользователя страница 20

 

Magstim BiStim² 

 

 

Operating Manual 

 

© The Magstim Company Limited 

17 

MOP02-EN-02 

5.3 

Voltage Selection and Fuse Rating 

The Magstim Bistim² system is preset at the factory to operate at 110V-120V or 220V-240V A.C., as 
appropriate. The following are the fuse ratings for both the master and slave units: 
 
 

Supply Voltage

 

Qty

 

Fuse Size

 

Rating

 

Voltage Selector 

Setting

 

110V-120V~ 

20 x 5mm 

T10AH, 250V 

115 

220V-240V~ 

20 x 5mm 

T5AH, 250V 

230 

 
 
T denotes timed or antisurge fuses. Fast acting fuses are not recommended. 
H denotes high breaking capacity (ceramic fuse). 
 

Fuse Replacement:

 The Magstim Bistim² must be disconnected from the mains by disconnecting the 

power cord from both of the 200² units. Insert the tip of a small blade screwdriver or similar tool into 
the slot below the arrow to lever the retaining lip enabling the fuse tray to be pulled out. Replace the 
blown  fuse(s)  with  one  of  the  correct  rating  (See  above  to  ensure  the  correct  fuse  is  replaced)  and 
locate and push home the tray ensuring the retaining lip clicks into place. 
 

Changing The Voltage Selector Setting:

 The Magstim Bistim² must be disconnected from the mains 

by  disconnecting  the  power  cord  from  both  of  the  200²  units.  Remove  the  fuse  tray  as  instructed 
above, rotate the voltage selector to the desired setting and replace the cover. A legend will indicate 
the voltage selected when the fuse cover is replaced. 
 

Note: Only replace fuses with those of the correct rating, and only operate the instrument with 
the voltage selector set at the appropriate voltage. Failure to do so may cause serious damage 
to the device(s). 

 

5.4 

Cleaning and Disinfecting

*

 

The  Magstim  Bistim²  system  and  accessories  cannot  be  sterilised;  therefore,  do  not  allow  them  to 
become contaminated with bodily fluids. They may be cleaned using a cloth moistened with isopropyl 
alcohol, however ensure that the Magstim Bistim² and accessories have dried thoroughly before use. 

It is the responsibility of the user to ensure the Magstim BiStim² is cleaned when necessary. 

Note:

  The  Magstim  Company  Ltd  recommends  that  a  solution  of  70%  isopropyl  alcohol  should  be 

used. 

 

5.5 

Servicing 

The  Magstim  Bistim²  contains  no  user  serviceable  parts.  Servicing  of  Magstim  Bistim²  system  and 
accessories  must  only  be  carried  out  by  The  Magstim  Company  Limited  or  one  of  its  authorised 
service centres. To arrange a return or for further information, contact the service department at The 
Magstim Company Limited (see Section 7.2 for contact details). 

 

5.6 

Device Lifetime 

The lifetime of the Magstim Bistim² system is defined as being five  years from the date of shipment. 
The Magstim Company Limited will support the products for the duration of their lifetime. 

Содержание BiStim2

Страница 1: ...January 2014 Magstim BiStim Operating Manual MOP02 EN Revision 02...

Страница 2: ...4 4 Operation 14 4 5 Independent BiStim Triggering IBT 15 Section 5 Maintenance 16 5 1 User Maintenance and Calibration 16 5 2 Technical Maintenance 16 5 3 Voltage Selection and Fuse Rating 17 5 4 Cle...

Страница 3: ...al service Claims for damages during shipment must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as th...

Страница 4: ...gstim Bistim is a non sterile reusable medical device intended for multi patient use The Magstim Bistim is not intended to come into contact with the patient as this achieved through use of a stimulat...

Страница 5: ...ent environment Note When the Bistim Connecting Module is not connected the Magstim 200 Units will operate as if they were a standalone device Please consult the Magstim 200 Operating Manual before us...

Страница 6: ...d to the Magstim Bistim This connection must be configured in compliance with IEC 60601 1 1 2000 or Clause 16 of IEC 60601 1 2005 Second edition IEC 60601 1 1 with the following interface voltage limi...

Страница 7: ...ng manual Stacking Limit Magstim 200 units may be stacked on top of each other but must not exceed the stacking limit specified in Section 6 of this operating manual If this limit is exceeded the stac...

Страница 8: ...trol system The only function button that may be used is the system stop button Figure 3 1 Detailed front panel The upper unit is considered the master unit The lower unit is considered the slave unit...

Страница 9: ...g connected correctly to the Magstim Bistim 3 1 6 Coil Temperature Indicators 6 These LED s represent the surface temperature of the coil The LED s illuminate sequentially as the coil temperature rise...

Страница 10: ...eases the level To toggle between the master power slave power and inter pulse spacing press the output control Each complete 360 rotation changes the master slave power output level by approximately...

Страница 11: ...MOP02 EN 02 Note It is normal for the pin indicated on the label to protrude from the coil socket on the master and slave stimulator units This is an intentional design feature Figure 3 2 Coil output...

Страница 12: ...xternal trigger input This is only functional on the master unit Please see Appendix B for more information 3 2 3 Foot Switch Socket 3 This socket is used to connect the pneumatic foot switch to the M...

Страница 13: ...unit prior to the unit being switched on 3 3 Accessories A range of stimulating coils are available for use with the Magstim range of products For more details please contact the Sales department Cont...

Страница 14: ...ating Manual Note The contacts in the coil connector Bistim connecting module and the stimulator coil output socket should be checked regularly for any signs of tarnishing or burning See Section 5 2 f...

Страница 15: ...key tighten the bolts so as to secure the coil holder to the right hand side of the master unit Do not over tighten the bolts as this may damage the cases of the units Figure 4 2 Attaching the coil ho...

Страница 16: ...l holder on the right hand side of the stimulator as instructed in Figure 4 3 Figure 4 4 Attaching the coil to the coil holder 5 Connect the stimulating coil to be used to the coil output socket on th...

Страница 17: ...ster unit UI to adjust the power level or the inter pulse spacing to the desired setting Rotating the control clockwise increases the power level rotating anti clockwise decreases the power level 4 De...

Страница 18: ...nnect the Magstim Bistim Note Ensure that the system is disarmed prior to disconnecting the stimulating coil and BiStim connecting module 4 5 Independent BiStim Triggering IBT When the system is set i...

Страница 19: ...ossible for localised heating to manifest itself in the connector which causes damage to the contacts Continued use in this condition will eventually cause the contacts to deteriorate and loose electr...

Страница 20: ...voltage selector to the desired setting and replace the cover A legend will indicate the voltage selected when the fuse cover is replaced Note Only replace fuses with those of the correct rating and...

Страница 21: ...erviceable life it should not be disposed of in general waste The Magstim Company Limited should be contacted see Section 7 2 for contact details for advice on its disposal in compliance with the appr...

Страница 22: ...ot comply with EMC Safety Standard EN 60601 1 2 including mobile phones Any interface cable which connects the Magstim Bistim unit to an external piece of equipment must be no more than 1 5m in length...

Страница 23: ...ife expectancy 2 x 10 6 discharges at 70 power level 8 x 10 5 discharges at 80 power level 4 x 10 5 discharges at 90 power level 2 x 10 5 discharges at 100 power level Isolated Interface Logic High Vo...

Страница 24: ...ering 6 5 Packing Instructions If for any reason it is necessary to return your Magstim Bistim care should be taken to ensure that the equipment is adequately packed to prevent transit damage Ideally...

Страница 25: ...imited Spring Gardens Whitland Carmarthenshire SA34 0HR Telephone 44 0 1994 240798 Fax 44 0 1994 240061 E mail info magstim com Website www magstim com 7 2 Servicing Enquiries Telephone 44 0 1994 2429...

Страница 26: ...gory C04 Invalid Power Identification C05 Temperature sensor circuit failure C06 Bad Serial command or received data checksum C07 Internal Software Error C08 Invalid Coil Temperature ID C09 Invalid te...

Страница 27: ...ck max difference E75 HVCAP over temperature E76 HV Transformer over temperature E77 Charge threshold failure E78 VREF check failure E79 HVCAP voltage comparison failure E80 Charging fault E81 HV over...

Страница 28: ...Bistim will only be triggered when the logic level being applied on the Tin or Tin pin is changed On the Tin pin triggering will occur when the logic level is changed from low to high On the Tin pin...

Страница 29: ...romagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Magstim Bistim can help prevent electromagnetic interference by maintaining a minimum distance b...

Страница 30: ...ital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT f...

Страница 31: ...tance in metres m Field Strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...

Отзывы: