MAGSTIM BiStim2 Скачать руководство пользователя страница 17

 

Magstim BiStim² 

 

 

Operating Manual 

 

© The Magstim Company Limited 

14 

MOP02-EN-02 

 

Figure 4.6: Diagram showing correct and incorrect ways to connect the coil to the stimulator 

 

6.  Connect the power leads to the rear of the master and slave units

*

 

4.4 

Operation 

1.  Set the power switch on the rear of the master and slave units to the on (I) position. The power 

indicators will flash indicating that the units are in standby

*

2.  Push  the  on/off  button  on  the  front  of  the  master  and  slave  units

*

.  The  power  indicators  will 

become  continuously  lit.  The  Magstim  Bistim²  system  will  start  an  initialisation  sequence  during 
which  the  armed  indicator 

will  flash  three  times  and  LED’s  in  the  user  interface  will  light 

sequentially.  

Once the initialisation sequence is complete, the coil indicator and disarmed indicator LED’s will 
be lit and the output display will read 30% on the first line, 30% on the  second line and 10ms on 
the third line. 

Note:

 

During the initialisation sequence, the operator must check that all LED’s are lit. If they do 

not,  the  items  must  not  be  used  and  should  be  returned  to  The  Magstim  Company  Limited  for 
servicing and repair. 

3.  Use the output control on the master unit UI to adjust the power level or the inter-pulse spacing to 

the  desired  setting

*

.  Rotating  the  control  clockwise  increases  the  power  level,  rotating  anti-

clockwise decreases the power level.

 

4.  Depress the output control to toggle the value being adjusted between the master power output, 

the slave power output and the inter-pulse spacing

*

.

 

The  inter-pulse  spacing  range  can  be  change  between  0-999ms  and  0-99.9ms,  when  adjusting 
the inter-pulse spacing, rotating the control knob on the clockwise several times until the display 
shows  the  maximum  inter-pulse  spacing  (999ms  or  99.9ms).  Press  and  hold  the  stop  button  on 
the  user  interface  and  rotate  the  control  knob  clockwise  at  the  same  time.    The  display  should 
revert to either 1.0ms or 10ms. Turn off the Bistim² system and then turn on again and the default 
inter-pulse time displayed should be in the new range (1.0ms or 10ms).

 

5.  Press  the  run(green)  button  on  the  master  unit  UI  to  arm  the  Magstim  Bistim²

*

.  The  armed 

indicators of both the master and the slave system should light, and once the system is charged, 
the ready indicators will also illuminate

*

Note:

  If  the  system  is  not  triggered  within  a  minute  of  being  armed,  it  will  revert  back  to  the 

disarmed state and will need to be re-armed before continuing. 

6.  Remove the stimulating coil from the coil holder* and position the coil over the desired target. 

Содержание BiStim2

Страница 1: ...January 2014 Magstim BiStim Operating Manual MOP02 EN Revision 02...

Страница 2: ...4 4 Operation 14 4 5 Independent BiStim Triggering IBT 15 Section 5 Maintenance 16 5 1 User Maintenance and Calibration 16 5 2 Technical Maintenance 16 5 3 Voltage Selection and Fuse Rating 17 5 4 Cle...

Страница 3: ...al service Claims for damages during shipment must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as th...

Страница 4: ...gstim Bistim is a non sterile reusable medical device intended for multi patient use The Magstim Bistim is not intended to come into contact with the patient as this achieved through use of a stimulat...

Страница 5: ...ent environment Note When the Bistim Connecting Module is not connected the Magstim 200 Units will operate as if they were a standalone device Please consult the Magstim 200 Operating Manual before us...

Страница 6: ...d to the Magstim Bistim This connection must be configured in compliance with IEC 60601 1 1 2000 or Clause 16 of IEC 60601 1 2005 Second edition IEC 60601 1 1 with the following interface voltage limi...

Страница 7: ...ng manual Stacking Limit Magstim 200 units may be stacked on top of each other but must not exceed the stacking limit specified in Section 6 of this operating manual If this limit is exceeded the stac...

Страница 8: ...trol system The only function button that may be used is the system stop button Figure 3 1 Detailed front panel The upper unit is considered the master unit The lower unit is considered the slave unit...

Страница 9: ...g connected correctly to the Magstim Bistim 3 1 6 Coil Temperature Indicators 6 These LED s represent the surface temperature of the coil The LED s illuminate sequentially as the coil temperature rise...

Страница 10: ...eases the level To toggle between the master power slave power and inter pulse spacing press the output control Each complete 360 rotation changes the master slave power output level by approximately...

Страница 11: ...MOP02 EN 02 Note It is normal for the pin indicated on the label to protrude from the coil socket on the master and slave stimulator units This is an intentional design feature Figure 3 2 Coil output...

Страница 12: ...xternal trigger input This is only functional on the master unit Please see Appendix B for more information 3 2 3 Foot Switch Socket 3 This socket is used to connect the pneumatic foot switch to the M...

Страница 13: ...unit prior to the unit being switched on 3 3 Accessories A range of stimulating coils are available for use with the Magstim range of products For more details please contact the Sales department Cont...

Страница 14: ...ating Manual Note The contacts in the coil connector Bistim connecting module and the stimulator coil output socket should be checked regularly for any signs of tarnishing or burning See Section 5 2 f...

Страница 15: ...key tighten the bolts so as to secure the coil holder to the right hand side of the master unit Do not over tighten the bolts as this may damage the cases of the units Figure 4 2 Attaching the coil ho...

Страница 16: ...l holder on the right hand side of the stimulator as instructed in Figure 4 3 Figure 4 4 Attaching the coil to the coil holder 5 Connect the stimulating coil to be used to the coil output socket on th...

Страница 17: ...ster unit UI to adjust the power level or the inter pulse spacing to the desired setting Rotating the control clockwise increases the power level rotating anti clockwise decreases the power level 4 De...

Страница 18: ...nnect the Magstim Bistim Note Ensure that the system is disarmed prior to disconnecting the stimulating coil and BiStim connecting module 4 5 Independent BiStim Triggering IBT When the system is set i...

Страница 19: ...ossible for localised heating to manifest itself in the connector which causes damage to the contacts Continued use in this condition will eventually cause the contacts to deteriorate and loose electr...

Страница 20: ...voltage selector to the desired setting and replace the cover A legend will indicate the voltage selected when the fuse cover is replaced Note Only replace fuses with those of the correct rating and...

Страница 21: ...erviceable life it should not be disposed of in general waste The Magstim Company Limited should be contacted see Section 7 2 for contact details for advice on its disposal in compliance with the appr...

Страница 22: ...ot comply with EMC Safety Standard EN 60601 1 2 including mobile phones Any interface cable which connects the Magstim Bistim unit to an external piece of equipment must be no more than 1 5m in length...

Страница 23: ...ife expectancy 2 x 10 6 discharges at 70 power level 8 x 10 5 discharges at 80 power level 4 x 10 5 discharges at 90 power level 2 x 10 5 discharges at 100 power level Isolated Interface Logic High Vo...

Страница 24: ...ering 6 5 Packing Instructions If for any reason it is necessary to return your Magstim Bistim care should be taken to ensure that the equipment is adequately packed to prevent transit damage Ideally...

Страница 25: ...imited Spring Gardens Whitland Carmarthenshire SA34 0HR Telephone 44 0 1994 240798 Fax 44 0 1994 240061 E mail info magstim com Website www magstim com 7 2 Servicing Enquiries Telephone 44 0 1994 2429...

Страница 26: ...gory C04 Invalid Power Identification C05 Temperature sensor circuit failure C06 Bad Serial command or received data checksum C07 Internal Software Error C08 Invalid Coil Temperature ID C09 Invalid te...

Страница 27: ...ck max difference E75 HVCAP over temperature E76 HV Transformer over temperature E77 Charge threshold failure E78 VREF check failure E79 HVCAP voltage comparison failure E80 Charging fault E81 HV over...

Страница 28: ...Bistim will only be triggered when the logic level being applied on the Tin or Tin pin is changed On the Tin pin triggering will occur when the logic level is changed from low to high On the Tin pin...

Страница 29: ...romagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Magstim Bistim can help prevent electromagnetic interference by maintaining a minimum distance b...

Страница 30: ...ital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT f...

Страница 31: ...tance in metres m Field Strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...

Отзывы: