3.9 Specifications
3.9.1 Performance
Supply Voltage
TD1 19.2 to 28.8 VDC
TD2 19.2 to 28.8 VDC
or 100 to 264 VAC
, 50–60 Hz
Power Consumption
TD1:
3.5 Watts at 24 VDC
4.5 Watts at 28.8 VDC
TD2:
4 Watts at 24 VDC
4.5 Watts at 28.8 VDC
5 Watts at 100 to 264 VAC
Power to Probe
Less than 1 Watt
Output Relay
TD1:
DPDT, 8 amp at 120 VAC, 250 VAC
(gold flashed)
8 amp at 28.8 VDC, 0.5 amp at 125 VDC
TD2:
DPDT, 8 amp at 120 VAC, 250 VAC
8 amp at 28.8 VDC, 0.5 amp at 125 VDC or
Hermetically Sealed DPDT 1 amp at 28 VDC, 0.2 amp at 125 VDC
Ambient Temperature
Electronics:
-40 to +158 °F (-40 to +70 °C)
Storage Temperature
Electronics:
-58 to +170 °F (-50 to +76 °C)
Operating Temperature
Sensor:
-100 to +400 °F (-73 to +204 °C)
➀
High Temperature Sensor:
-100 to +850 °F (-73 to +454 °C)
Response Time
1–10 seconds (typical – dependent upon sensor type, application,
and set point adjustment)
Set Point Range
Water:
0.01 to 5.0 FPS (0.003 to 1.5 m/s) (spherical tip and twin tip sensors)
0.01 to 1.0 FPS (0.003 to 0.3 m/s) (HTHP, Hastelloy, Monel sensors)
Air:
0.1 to 500 SFPS (0.03 to 150 Nm/s)
Time Delay (TD2 only)
0–100 seconds adjustable (time delay in addition to sensor response)
Repeatability
<1% at constant temperature
Enclosure Material
Cast aluminum A356 containing less than 0.2% copper
316 stainless steel or 304 stainless steel
SIL
Safe Failure Fraction (SFF) TD1=69.3% TD2=73%
3.9.2 Probe
Probe Materials of Construction
Twin Tip:
All wetted parts of 316/316L stainless steel, Hastelloy C, or Monel
HTHP Sensor:
316/316L stainless steel, Hastelloy C
Mini Sensor:
316/316L stainless steel
Spherical Tip:
316/316L stainless steel
Low Flow Body:
316/316L stainless steel
Process Connection
Refer to part number construction
Probe Insertion Length:
Spherical Tip Probe, TXA, TXB
Available in lengths from 2 to 130" in 1" increments
Twin Tip Probe, TXC, TXD
(5 to 330 cm in 1 cm increments)
Mini Sensor, TXM
Available insertion lengths 1 to 130" (3 to 330 cm)
High Temperature Sensor, TXH
Available in lengths from 2 to 36" in 1" increments
➁
(5 to 91 cm in 1 cm increments)
Low Flow Body, TEL
1
⁄
4
" and
1
⁄
2
" NPT and G (BSP) threads.
Cable Length
500 feet (150 meters) maximum
➀
Use a probe with heat extension or remote electronics for process temperatures greater than +250 °F (+120 °C)
➁
Longer lengths available. Consult factory.