51-638 Echotel Model 338 Two-Wire Ultrasonic Transmitter
38
2.6.2 Transmitter
Supply voltage
12 to 36 VDC
Power consumption
48 to 720 mW
Output signal
4-20 mA isolated,
600 ohm load
Diagnostic alarm
3.6, 22 mA, or Hold
User Interface:
Keypad
4-button menu-driven data
entry
Display
Removable 6 digit custom
LCD module with dual
function bar graph
LED status indication
Echo strength and power
Ambient temperature
-22° to +140° F
(-30° to +60° C)
LCD module temperature
-13° to +140° F
(-25° to +60° C)
Enclosure
Plastic (PBT)
NEMA 6 (IP 67)
Cable entry
Two –
1
⁄
2
" NPT and
Two – PG16
2.6.3 Transducer
Maximum range
20 feet (6 meters)
Dead zone
10 inches (250 mm)
Frequency
80 kHz
Process connection
2 inch NPT
(flanges optionally available)
Process temperature
-22° to +195° F
(-30° to +90° C)
Process pressure
Atmospheric to 44 psi (3 bar)
Ultrasonic beam angle
5° conical
Temperature compensation
Automatic over the operating
temperature range of the
transducer
Transducer
Polypropylene NEMA 6P
(IP 68)
These units have been tested to EN 50081-2
and EN 50082-2 and are in compliance with
the EMC Directive 89/336/EEC.