Magnetrol Jupiter JM4 Скачать руководство пользователя страница 27

52 

ORI-46.650 Jupiter Magnetostrictive Transmitters

ORI-46.650 Jupiter Magnetostrictive Transmitters 

53

2.5 

Specifications

2.5.1  Functional/Physical 

 

System Design

  Measurement Principle

Magnetostriction-based mechanical response signal

Input

  Measured Variable

Level, response signal time of flight

 Span

6 inches to 400 inches (15 cm to 999 cm)

Output

  Type

4 to 20 mA with HART: 3.8 mA to 20.5 mA useable (per NAMUR NE43)
Foundation fieldbus

TM

: H1 (ITK Ver. 6.1.1)

  Resolution                   Analog:

0.003 mA

                            Digital Display:

1 mm

  Loop Resistance

591 ohms @ 24 VDC and 22 mA

  Diagnostic Alarm

Selectable: 3.6 mA, 22 mA (meets requirements of NAMUR NE 43), or HOLD last output

 Damping

Adjustable 0-10 seconds

User Interface

  Keypad

4-button menu-driven data entry

  Display

Graphic liquid crystal display

  Digital Communication

HART Version 7 —with Field Communicator, Foundation fieldbus

TM

DTM (PACT

ware

TM

), AMS, FDT, EDDL

  Menu Languages

Transmitter LCD: English, French, German, Spanish, Russian, Portuguese
HART DD: English, French, German, Spanish, Russian, Chinese, Portuguese

Foundation fieldbus

TM

 Host System: English

Power 

(at transmitter terminals)

HART: General Purpose (Weather proof)/Intrinsically Safe/Explosion-proof:
11 VDC minimum under certain conditions (refer to Section 2.5.5)
FOUNDATION fieldbus™: 9 to 17.5 VDC

FISCO, FNICO, Explosion Proof, General Purpose, and Weather Proof

Housing

 Material

IP67/die cast aluminum A413 (<0.6% copper); optional 316 stainless steel

  Net/Gross Weight

Aluminum: 4.5 lbs. (2.0 kg)
Stainless Steel: 10.0 lbs. (4.50 kg)

  Overall Dimensions

H 8.34” (212 mm) x W 4.03” (102 mm) x D 7.56” (192 mm)

  Cable Entry

1⁄2” NPT or M20

  SIL 2 Hardware (Safety Integrity Level)

Safe Failure Fraction = 93.1% for Single Float version,
91.9% for Dual Float version (HART only)
Functional Safety to SIL 2 as 1oo1 in accordance with IEC 61508
(Full FMEDA report available upon request)

Environment

Ambient Temperature Range         Transmitter:

-40°F to +176°F (-40°C to +80°C)

                                                                  LCD:

-5°F to +176°F (-20°C to +80°C)

Process Temperature              External Mount:

-320°F to +850°F (-195°C to +454°C)

                                                Direct Insertion:

-320°F to +500°F (-195°C to +260°C)
-320°F to +800°F (-195°C to +427°C)

Storage Temperature

-50°F to +185°F (-46°C to +85°C)

Process Pressure (Direct Insertion)

Vacuum to 3000 psig (207bar)

      

See page 41 for Ambient Temperature Range based off protection method

      

See Special Conditions of Use Item 2 (page 42)

2

2

1

1

2.5.1  Functional/Physical 

 

  Humidity

0 to 99%, non-condensing

  Electromagnetic Compatibility

Meets CE requirement (EN 61326) and NAMUR NE 21

  Surge Protection

Meets CE EN 61326 (1000V)

 Shock/Vibration

ANSI/ISA-S71.03 Class SA1 (Shock); ANSI/ISA-S71.03 Class VC2 

(Vibration)

Performance

  Linearity

0.030 in. or  0.01% of probe length, whichever is greater

  Accuracy

±0.01% full scale or 0.05 in, whichever is greater

 Resolution

.014” (.35 mm)

  Repeatability

±0.005% of full span or 0.014 in, whichever is greater

  Response Time

1 second

  Initialization Time

Less than 10 seconds

  Ambient Temperature Effect

Approx.  ±0.02% of probe length/degree C

Foundation fieldbus

TM

  ITK Version

6.1.1

  H1 Device Class

Link Master (LAS)—selectable ON/OFF

  H1 Profile Class

31PS, 32L

  Function Blocks

(6) Al, (2) Transducer, (1) Resource, (1) Arithmetic, (1) Input Selector,
(1) Signal Characterizer, (2) PID, (1) Integrator

  Quiescent Current

15 mA

  Execution Time

15 msec (30 msec PID, Signal Characterizer Block)

2.5.2  Probe Selection Guide   

JM4 Probe

Temperature Range

Pressure Range (Direct Insertion)

Standard

-50

°

F to 450

°

F  (-46

°

C to 232

°

C)

Vacuum to 3000 psig (207 bar)

High Temperature

-50

°

F to 800

°

F (-46

°

C to 427

°

C)

Vacuum to 3000 psig (207 bar)

Cryogenic

-320

°

F to 158

°

F (-196

°

C to 70

°

C) Vacuum to 3000 psig (207 bar)

2.5.3  Probe Specifications 

 

Probe Diameter

0.625 inches (1.59 cm)

Process Connection

3/4” NPT, 2” NPT, 1” BSP, 2” BSP

ANSI or DIN Flanges

Materials

Standard: 316 SS, nickel, Teflon

®

, brass, silicone 

High Temperature: 316 SS, ceramic, brass, nickel, mica

Cryogenic: 316 SS, silicone, nickel, Teflon

®

, brass

Содержание Jupiter JM4

Страница 1: ...Model JM4 Magnetostrictive Level Transmitter HART Installation and Operating Manual Software v1 x...

Страница 2: ...ation labor claims direct or consequential damage or expense arising from the installation or use of equipment There are no other warranties expressed or implied except special written warranties cove...

Страница 3: ...nents or connector pins Make sure that all electrical connections are completely made and none are partial or floating Ground all equipment to a reli able earth ground 1 1 4 Site Preparation Power Sup...

Страница 4: ...eft to right in Figure 1 4 top mount top mount offset bottom mount offset All of these configurations are available with one or two magnetic floats two floats are necessary to measure total AND interf...

Страница 5: ...th an insulation blanket secure the blanket over the chamber then be sure to place the transmitter in the smaller of the two grooves leaving the larger groove for visual indication Figure 1 6b 1 2 7 V...

Страница 6: ...e to the terminal 6 Replace and tighten the cover to the transmitter wiring compartment before applying power 1 4 2 Intrinsically Safe An Intrinsically Safe IS installation potentially has flammable m...

Страница 7: ...avigation and data entry The Model JM4 user interface is hierarchical in nature best de scribed as a tree structure Each level in the tree contains one or more items Items are either menu labels or pa...

Страница 8: ...The user password can be changed to any numerical value up to 59999 When the transmitter is programmed for password protection a password is required whenever configuration values are changed User Pa...

Страница 9: ...AIN MENU will result in an LCD presentation as shown at left The small down arrow shown at the right hand side of the screen is the indication that more items are available below and can be accessed b...

Страница 10: ...rs Upper Thickness Units Inches Feet Millimeters Centimeters Meters Device Setup Main Menu Home Screen Product Name read only Orion S N read only DIGITAL BOARD read only ANALOG BOARD read only Physica...

Страница 11: ...l Display Config Advanced Config Factory Config Blocking Distance 0 to 50 feet 0 m to 15 m Minimum Separation 0 to 28 inches 0 cm to 71 cm Zero Offset 50 to 50 feet 15 m to 15 m Level Trim 2 to 2 feet...

Страница 12: ...JM4 software Device Descriptions Refer to your HART Communicator Manual for update instructions 1 6 1 Connections A HART communicator can be operated from a remote loca tion by connecting it to a rem...

Страница 13: ...6 4 HART Menu Model JM4 NOTE Numbered menu boxes below correspond to numbered tables on Section 1 6 5 1 PV 2 PV Analog Output 3 PV Output 4 Device Setup 5 Setup Wizard 6 Diagnostics 7 Measured Values...

Страница 14: ...his value corresponds to the level of the upper surface 2 PV Analog Output Digital representation that tracks the Analog Output Number 1 under normal operating modes Analog Output Value Value that tra...

Страница 15: ...ion or calibration has been changed by a host application or from a local operator interface 9 Device ID Uniquely identifies the Field Device when combined with the Manufacturer Identification and Dev...

Страница 16: ...n operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived Ifc 4mA Set Point Thk 4mA Set Point Vol 4mA Se...

Страница 17: ...t to match the polarity of the first portion of the sine wave signal of a direct insertion float or the first portion of the M shaped signal of an external mount probe Ifc Lvl Polarity 1 6 5 HART Menu...

Страница 18: ...es only 1 6 5 HART Menu Items cont 19 Display Description 1 Present Temperature Present temperature measured in the electronics compartment 2 Max Temperature Maximum temperature measured in the electr...

Страница 19: ...e mode this value corresponds to the strength of the upper level signal 6 Ifc Echo Strength Relative strength of the interface level signal 7 Analog Output Allows for configuration of the 4 20mA Analo...

Страница 20: ...valuable proactive method of troubleshooting The device not only tells the user what wrong but also and more impor tantly offers suggestions on how to solve the problem All of this information can be...

Страница 21: ...Specification and Maintenance Required The FOUNDATION fieldbus transmitter version of the Model JM4 was implemented according to the Field Diagnostics Pro file which is consistent with the objectives...

Страница 22: ...upper level threshold Check echo curve for noise levels Echo Rejection may be required Consult Factory Swipe probe with pocket magnet to eliminate possibility of residual magnetism High Noise Ifc Thre...

Страница 23: ...rtical Flat factory default Vessel Type Vertical Elliptical Vertical Spherical Vertical Conical Custom Table Rectangular Horizontal Flat Horizontal Elliptical Horizontal Spherical or Spherical Note Ve...

Страница 24: ...roubleshooting this parameter gives the user the ability to reset the Model JM4 transmitter configuration Unique to the Model JM4 transmitter is the ability for Orion In struments to fully pre configu...

Страница 25: ...EC60079 11 2011 IEC60079 15 2010 IEC60079 31 2008 SPECIAL CONDITIONS OF USE 1 The enclosure contains aluminum and is considered to present a potential risk of ignition by impact or friction Care must...

Страница 26: ...NSTALLATION REFERENCE ALL APPLICABLE NOTES FROM PAGE 2 99 5074 001 Ui Vmax 24V Ii Imax 280mA Pi 1 93W or Any approved termination on with R 90 100 C 0 2 2 F Any FM CSA Approved Intrinsically Safe Asso...

Страница 27: ...upon request Environment Ambient Temperature Range Transmitter 40 F to 176 F 40 C to 80 C LCD 5 F to 176 F 20 C to 80 C Process Temperature External Mount 320 F to 850 F 195 C to 454 C Direct Inserti...

Страница 28: ...tion 75 NPT High Temp Cryo Direct Insertion 1 BSP 16 92 429 7 Direct Insertion 2 BSP 17 17 436 Direct Insertion 2 NPT 2 5 4 Physical Specifications Transmitter inches mm 18 92 480 5 22 92 582 1 Direct...

Страница 29: ...SIC MODEL NUMBER JM4 Jupiter 4th Generation Magnetostrictive Level Transmitter 4 POWER 5 24 VDC Two Wire 5 SIGNAL OUTPUT 1 4 20 mA with HART 2 Foundation Fieldbus Communications 6 SAFETY OPTIONS 1 SIL...

Страница 30: ...ct Insertion float provided or unit is External Mount Jupiter 99 Custom Direct Insertion float not listed in standard offering below One Float Total Level Only Digit 10 1 AA 316L SS MIN SG 0 83 2 00 O...

Страница 31: ...r External Mount Probe 0 0 0 2 1 3 2 4 5 6 7 8 9 10 11 12 13 14 15 8 INSTALLATION CONSIDERATIONS 0 No mounting clamps required Chamber Clamp Sizes for MLI Mount Without High Temperature Insulation 1 2...

Страница 32: ...closure Fault A defect or failure in a circuit The current mA value unit defaults to 3 6 22 or Hold when a diagnostic condition occurs Feedthrough A small connecting cavity between the main housing co...

Страница 33: ...to user s meter DCS input etc Two Wire An electrical instrument design that uses one set of wires to provide both the supply power and process measurement signal The process measurement is achieved b...

Страница 34: ...probe 4mA or 0 20mA or 100 Is moderate to heavy vibration present Is turbulence or flow present Additional application details that could impact performance Will existing chamber or stilling well be u...

Страница 35: ...red trademarks of Magnetrol International Incorporated HART is a registered trademark of the HART Communication Foundation PACTware is trademark of PACTware Consortium CSA logotype is a registered tra...

Отзывы: