Magnetrol Jupiter JM4 Скачать руководство пользователя страница 21

40 

ORI-46.650 Jupiter Magnetostrictive Transmitters

ORI-46.650 Jupiter Magnetostrictive Transmitters 

41

2.3.2  Troubleshooting Table   

Problem

Solution

Blank display

Ensure local Keypad / LCD  is properly installed.  

Remove power and reapply power to the unit. 

Check to see if LED on module is illuminated.

Check voltage at terminal board.

If jumper is in place under display, remove jumper.

Transmitter does not track level            (External Mount)

(Direct Insertion)

Check echo curve for noise preventing tracking of level.

Remove transmitter and probe from piping column and 

test with re-alignment magnet. Run magnet from bottom 

to top of probe. 

Check zero and span calibration. If no change in output, 

consult the factory.

Float inside the level gauge is moving slowly or not at all.

Ensure that the magnetic level indicator is plumb.

The process fluid being measured may be too viscous 

and heat tracing may be required to make the material 

more fluid.

The specific gravity of the process fluid and float weight 

may need to be reverified.

The liquid being measured may contain magnetic 

particles collecting on the magnetic section of the float 

causing drag. If this happens magnetic trap assemblies 

can be purchased from the factory.

Visual inspection of the float may be required to see if 

the float has collapsed.

LEVEL, % OUTPUT, and LOOP values are all 

inaccurate.

Wipe probe with external magnet.

Confirm configuration settings.

Consult factory.

LEVEL, % OUTPUT, and LOOP values fluctuate.

Check echo curve for noise levels that could be affecting 

level reading.

Turbulence, increase damping factor until readings 

stabilize.

Level reading on display is correct, but loop value is 

stuck at 4 mA.

Set poll address to zero.

As mentioned above, the indicators can be assignable (via the 

a DTM or host system) by the user to any (or none) of the 

NAMUR recommended Status Signal categories: Failure, Func-

tion Check, Out of Specification, and Maintenance Required.
The FOUNDATION fieldbus transmitter version of the Model 

JM4 was implemented according to the Field Diagnostics Pro-

file, which is consistent with the objectives of NE 107.
In the FOUNDATION fieldbus version, diagnostic indicators 

can be mapped to multiple categories, an example is shown in 

the diagram at left.
In the FOUNDATION fieldbus version, diagnostic indicators 

can be mapped to multiple categories, an example is shown in 

the diagram at left.
In this example, “Calibration Required” is mapped to both the 

Out of Specification and Maintenance Required status signals, 

and the diagnostic indicator named “High Temperature” is 

mapped to none of the signals.
Indicators that are mapped to the Failure category will normally 

result in a current loop alarm output. The alarm state for HART 

transmitters is configurable as high (22 mA), Low (3.6 mA), or 

Hold (last value).
Users will not have the ability to unassign certain indicators 

from the Failure signal category as the Model JM4 user interfac-

es will prohibit or reject such re-assignment entries). This is to 

ensure that current loop alarms are asserted in situations where 

the device is not able to provide measurements due to critical 

failures. (For example, if the alarm selection has not been set to 

Hold, or a fixed current mode is in effect.)
A default mapping of all diagnostic indicators will be applied 

initially, and can be re-applied through use of a reset function.
Refer to the table below for a complete listing of the Model JM4 

diagnostic indicators, along with their explanations, default cate-

gories, and recommended remedies.

NOTES: 1) The remedies shown in this table can also be seen on the 

transmitter LCD by viewing the present status screen when 

the device is in a diagnostic condition.

 

2) Those indicators showing failure as the default result in an  

alarm condition.

Analog Output Error

Failure

Function

Check

Out of

Specification

Maintenance

Required

Diagnostic Indicators

Calibration

Required

High 

Temperature

Echo Lost

Dry

Probe

Fig. 2-2. Namur NE 107 Status Signals - Venn 

Diagram

Содержание Jupiter JM4

Страница 1: ...Model JM4 Magnetostrictive Level Transmitter HART Installation and Operating Manual Software v1 x...

Страница 2: ...ation labor claims direct or consequential damage or expense arising from the installation or use of equipment There are no other warranties expressed or implied except special written warranties cove...

Страница 3: ...nents or connector pins Make sure that all electrical connections are completely made and none are partial or floating Ground all equipment to a reli able earth ground 1 1 4 Site Preparation Power Sup...

Страница 4: ...eft to right in Figure 1 4 top mount top mount offset bottom mount offset All of these configurations are available with one or two magnetic floats two floats are necessary to measure total AND interf...

Страница 5: ...th an insulation blanket secure the blanket over the chamber then be sure to place the transmitter in the smaller of the two grooves leaving the larger groove for visual indication Figure 1 6b 1 2 7 V...

Страница 6: ...e to the terminal 6 Replace and tighten the cover to the transmitter wiring compartment before applying power 1 4 2 Intrinsically Safe An Intrinsically Safe IS installation potentially has flammable m...

Страница 7: ...avigation and data entry The Model JM4 user interface is hierarchical in nature best de scribed as a tree structure Each level in the tree contains one or more items Items are either menu labels or pa...

Страница 8: ...The user password can be changed to any numerical value up to 59999 When the transmitter is programmed for password protection a password is required whenever configuration values are changed User Pa...

Страница 9: ...AIN MENU will result in an LCD presentation as shown at left The small down arrow shown at the right hand side of the screen is the indication that more items are available below and can be accessed b...

Страница 10: ...rs Upper Thickness Units Inches Feet Millimeters Centimeters Meters Device Setup Main Menu Home Screen Product Name read only Orion S N read only DIGITAL BOARD read only ANALOG BOARD read only Physica...

Страница 11: ...l Display Config Advanced Config Factory Config Blocking Distance 0 to 50 feet 0 m to 15 m Minimum Separation 0 to 28 inches 0 cm to 71 cm Zero Offset 50 to 50 feet 15 m to 15 m Level Trim 2 to 2 feet...

Страница 12: ...JM4 software Device Descriptions Refer to your HART Communicator Manual for update instructions 1 6 1 Connections A HART communicator can be operated from a remote loca tion by connecting it to a rem...

Страница 13: ...6 4 HART Menu Model JM4 NOTE Numbered menu boxes below correspond to numbered tables on Section 1 6 5 1 PV 2 PV Analog Output 3 PV Output 4 Device Setup 5 Setup Wizard 6 Diagnostics 7 Measured Values...

Страница 14: ...his value corresponds to the level of the upper surface 2 PV Analog Output Digital representation that tracks the Analog Output Number 1 under normal operating modes Analog Output Value Value that tra...

Страница 15: ...ion or calibration has been changed by a host application or from a local operator interface 9 Device ID Uniquely identifies the Field Device when combined with the Manufacturer Identification and Dev...

Страница 16: ...n operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived Ifc 4mA Set Point Thk 4mA Set Point Vol 4mA Se...

Страница 17: ...t to match the polarity of the first portion of the sine wave signal of a direct insertion float or the first portion of the M shaped signal of an external mount probe Ifc Lvl Polarity 1 6 5 HART Menu...

Страница 18: ...es only 1 6 5 HART Menu Items cont 19 Display Description 1 Present Temperature Present temperature measured in the electronics compartment 2 Max Temperature Maximum temperature measured in the electr...

Страница 19: ...e mode this value corresponds to the strength of the upper level signal 6 Ifc Echo Strength Relative strength of the interface level signal 7 Analog Output Allows for configuration of the 4 20mA Analo...

Страница 20: ...valuable proactive method of troubleshooting The device not only tells the user what wrong but also and more impor tantly offers suggestions on how to solve the problem All of this information can be...

Страница 21: ...Specification and Maintenance Required The FOUNDATION fieldbus transmitter version of the Model JM4 was implemented according to the Field Diagnostics Pro file which is consistent with the objectives...

Страница 22: ...upper level threshold Check echo curve for noise levels Echo Rejection may be required Consult Factory Swipe probe with pocket magnet to eliminate possibility of residual magnetism High Noise Ifc Thre...

Страница 23: ...rtical Flat factory default Vessel Type Vertical Elliptical Vertical Spherical Vertical Conical Custom Table Rectangular Horizontal Flat Horizontal Elliptical Horizontal Spherical or Spherical Note Ve...

Страница 24: ...roubleshooting this parameter gives the user the ability to reset the Model JM4 transmitter configuration Unique to the Model JM4 transmitter is the ability for Orion In struments to fully pre configu...

Страница 25: ...EC60079 11 2011 IEC60079 15 2010 IEC60079 31 2008 SPECIAL CONDITIONS OF USE 1 The enclosure contains aluminum and is considered to present a potential risk of ignition by impact or friction Care must...

Страница 26: ...NSTALLATION REFERENCE ALL APPLICABLE NOTES FROM PAGE 2 99 5074 001 Ui Vmax 24V Ii Imax 280mA Pi 1 93W or Any approved termination on with R 90 100 C 0 2 2 F Any FM CSA Approved Intrinsically Safe Asso...

Страница 27: ...upon request Environment Ambient Temperature Range Transmitter 40 F to 176 F 40 C to 80 C LCD 5 F to 176 F 20 C to 80 C Process Temperature External Mount 320 F to 850 F 195 C to 454 C Direct Inserti...

Страница 28: ...tion 75 NPT High Temp Cryo Direct Insertion 1 BSP 16 92 429 7 Direct Insertion 2 BSP 17 17 436 Direct Insertion 2 NPT 2 5 4 Physical Specifications Transmitter inches mm 18 92 480 5 22 92 582 1 Direct...

Страница 29: ...SIC MODEL NUMBER JM4 Jupiter 4th Generation Magnetostrictive Level Transmitter 4 POWER 5 24 VDC Two Wire 5 SIGNAL OUTPUT 1 4 20 mA with HART 2 Foundation Fieldbus Communications 6 SAFETY OPTIONS 1 SIL...

Страница 30: ...ct Insertion float provided or unit is External Mount Jupiter 99 Custom Direct Insertion float not listed in standard offering below One Float Total Level Only Digit 10 1 AA 316L SS MIN SG 0 83 2 00 O...

Страница 31: ...r External Mount Probe 0 0 0 2 1 3 2 4 5 6 7 8 9 10 11 12 13 14 15 8 INSTALLATION CONSIDERATIONS 0 No mounting clamps required Chamber Clamp Sizes for MLI Mount Without High Temperature Insulation 1 2...

Страница 32: ...closure Fault A defect or failure in a circuit The current mA value unit defaults to 3 6 22 or Hold when a diagnostic condition occurs Feedthrough A small connecting cavity between the main housing co...

Страница 33: ...to user s meter DCS input etc Two Wire An electrical instrument design that uses one set of wires to provide both the supply power and process measurement signal The process measurement is achieved b...

Страница 34: ...probe 4mA or 0 20mA or 100 Is moderate to heavy vibration present Is turbulence or flow present Additional application details that could impact performance Will existing chamber or stilling well be u...

Страница 35: ...red trademarks of Magnetrol International Incorporated HART is a registered trademark of the HART Communication Foundation PACTware is trademark of PACTware Consortium CSA logotype is a registered tra...

Отзывы: