7
Programming relay switch-off point (the level where relay becomes de-energised)
Place the Echotel at a distance to the target corresponding to the required switch-off point. (Do not forget to check valid Echo!)
Action
LED indication
1) Place magnet to symbol “
A
”
=Programming mode
2) Place magnet to symbol “
B
” and
=Programming in progress
3) Hold magnet to symbol “
B
”
=Programming in progress
4) Keep holding magnet to symbol “
B
”
=Programming in progress
5) Remove magnet when all LEDs are off
=End of Programming
Use level in tank or a fix target, i.e. the wall
Please note that the smallest switch-differential achievable with magnet programming is 20 mm.
“Error indication” by the current output (Check valid echo as above)
Action
LED indication
1) Place magnet to the symbol “
A
”
= Transmitter in programming mode
2) Place magnet to the symbol “
B
”
repeatedly
to select the required error indication mode
= Hold last value
= 3.6 mA
= 22 mA
3) Place magnet to the symbol “
A
”
= Programming completed
“Damping” (Check valid echo as above)
Action
LED indication
1) Place magnet to the symbol “
B
”
= Transmitter in programming mode
2) Place magnet to the symbol “
A
”
repeatedly
to select the required damping
= 10 sec
= 30 sec
= 60 sec
3) Place magnet to the symbol “
B
”
= Programming completed
Reset (to factory default)
Action
LED indication
1) Place magnet to the symbol “
B
”
= Programming mode
2) Place magnet to the symbol “
A
and
=
Reset in progress
3) Hold magnet to the symbol “
A
”
= Reset in progress
4) Remove magnet when all LEDs are off
= End of programming
Error indications during programming (by LEDs)
Action
LED status = error indicated
Correction
1) Attempted programming
= blinking twice = No Echo
Find valid echo
2) Attempted programming
= blinking three times = access denied (access code
active)
With use of 375 - 0SAP -100 see Chapter 5.2
(P99)
3) Attempted programming
= blinking four times = Echotel not in LEV meas. mode
With use of 375 - 0SAP -100 see Chapter 5.2
(P01)
4) Programming of the relay
=
blinking alternately = switch-differential too small
Set switch-differential greater than 20 mm