5.2.5 QUICKSET
Recommended as a simple and fast way to start up Echotel.
QUICKSET programming is aided by 8 screens to set the 8 basic
parameters of the device if the required application is uncomplicated level
metering, recommended for liquids only.
The instructions of this programming mode are also to be found, below
the screw cover, on the front panel of the Echotel.
The DEFAULT of the Current output, Display and Bargraph
is LEVEL.
This can be modified only in the Full Parameter Access mode see 5.1
(P01).
Keys
Function
ENTER
E
+ DOWN
(press for 3 seconds)
Enter or exit QUICKSET programming mode
UP
/ DOWN
Increase/decrease blinking digit or scroll up/down
NEXT
Move left with the blinking digit
UP
+ DOWN
"GET LEVEL" - display actual level measured by Echotel
ENTER
E
Save value on the screen and move to the next screen
NEXT
+ UP
Quit QUICKSET programming mode without saving the modifications
NEXT
+ DOWN
Display Factory Default of the relevant screen
See next page for details.
Screens
Actions
AP
plication
xx
= select “EU” (European) for metric or “US” for US engineering units
(Use UP
/ DOWN
keys)
yy= indicating “Li” for liquids or “So” for solids level measurement (can not be changed).
DEFAULT: EU
H
=
xxxx
maximum measuring distance
– Distance between transducer face and tank/silo bottom
Manual: set value
(Use UP
/ DOWN
/
NEXT
keys)
and save it
(by
ENTER
E
)
Automatic: use the “GET LEVEL” function
(UP
+ DOWN
)
to obtain actual measured value with level in tank or a fix
target, i.e. wall. (“GET LEVEL” functions only if ECHO LED is lit) and save it as above.
DEFAULT: maximum measuring distance [m], see Technical Data Table
4
mA
xxxx
–
level value
assigned to 4 mA current output
Manual: set level value
(by UP
/ DOWN
/
NEXT
keys)
and save it
(by ENTER
E
)
Automatic: use the “GET LEVEL” function
(UP
+ DOWN
)
to display the actual measured value with level in tank or a fix
target, i.e. wall. (“GET LEVEL” functions only if ECHO LED is lit) and save it as above.
DEFAULT: 0 m (0%, Empty tank)
20
mA
xxxx
–
level value
assigned to 20 mA current output
Manual: set level value
(Use UP
/ DOWN
/
NEXT
keys)
and save it (by
ENTER
E
)
Automatic: use the “GET LEVEL” function
(UP
+ DOWN
)
to obtain actual measured value with level in tank or a fix
target, i.e. wall. (“GET LEVEL” functions only if ECHO LED is lit) and save it as above.
DEFAULT: max. level = max. measuring distance – dead band [m] (100%, Full tank) (See Technical Data Table)
Er
ror indication by the current output
– select “Hold”, “3.6” mA or “22” mA
(by UP
/ DOWN
key)
and save it as above.
DEFAULT: hold last value
d
amping
t
ime
:
select required damping time
(by UP
/ DOWN
key)
and save it as above.
DEFAULT: 60 sec for liquids, 300 sec for solids
r
elay
E
nergised
xxxx
:
level
of the relay energised state
If the value exceeds this programmed value the relay will be energised
r
elay
d
e-
e
nergised
xxxx
:
level
of the relay de-energised state
If the value sinks below this programmed value the relay will be deenergised
Note: Current output can also be programmed for inverted operation: 4 mA= 100% (Full), 20 mA= 0% (Empty)