LeadCare Ultra
®
Blood Lead Testing System User’s Guide
8-10
8.5 LIS Interface
The Ultra User Application provides the capability to transmit approved patient test results using the
HL7 v2.5.1 standard. The LIS initiates the connection over TCP/IP on a specified port. Valid ports range
from 1 – 65535. Once the connection has been established, the Ultra User Application will transmit all
approved results per the format detailed in the document titled Connectivity to LIS.
NOTE:
Refer to the Connectivity to LIS document for LeadCare Ultra which
describes how data is transmitted between the Ultra application and an
external system such as LIS. If you need this document, please contact
Product Support.
To configure the parameters in the Ultra User App go to
Tools > LIS > Settings
•
In the LIS Configuration Box that opens (Figure 8-12), check the Enable LIS Interface option.
•
Enter the TCP port to be used for transfer of results. Valid ports range from1 – 65535.
•
Click the
Save
button.
NOTE: The Ultra User Application MUST BE RESTARTED
if one of the
parameters has been modified.
NOTE:
Only Admin accounts have the privileges to modify the LIS settings. All User
accounts can view the settings.
Figure 8-12: LIS Configuration
Содержание LeadCare Ultra
Страница 16: ...LeadCare Ultra Blood Lead Testing System User s Guide 2 6 This page intentionally left blank ...
Страница 62: ...LeadCare Ultra Blood Lead Testing System User s Guide 6 20 This page intentionally left blank ...
Страница 92: ...LeadCare Ultra Blood Lead Testing System User s Guide 11 2 This page intentionally left blank ...
Страница 96: ...LeadCare Ultra Blood Lead Testing System User s Guide 13 2 This page intentionally left blank ...
Страница 102: ...LeadCare Ultra Blood Lead Testing System User s Guide A 4 This page intentionally left blank ...
Страница 103: ...LeadCare Ultra Blood Lead Testing System User s Guide B 1 ...
Страница 104: ...LeadCare Ultra Blood Lead Testing System User s Guide B 2 ...
Страница 116: ...LeadCare Ultra Blood Lead Testing System User s Guide B 12 This page intentionally left blank ...
Страница 118: ...LeadCare Ultra Blood Lead Testing System User s Guide C 2 This page intentionally left blank ...