SIGMA LASER STENT AND TUBE CUTTER
990-401
19
12.
Attach the cutting bushing to the cutting clamp using the M2.5 pan head and loosely tighten.
13.
Bolt the cutting clamp to the cutting block holder using 4 8-32 cap head screws. Insure that the
Bellville washer is in place use the tube to be processed to center the clamp and provide
clearance for the Bellville washer by placing the tube in the 5C collet and in the cutting collet.
14.
While tightening the 4 bolts in a alternate bolt pattern have
the X axis moving in a positive and negative motion, or by
manually moving the tubing in a positive and negative
motion.
NOTE:
This movement while tightening the 4 cap head
screws will insure that the cutting collet, clamp, and clamp
bushing are all concentric.
15.
Tighten the M2.5 pan head screw in the bushing.
16.
Push the air supply tube into the air fitting mounted on the
clamp. If required the height of the tooling block can be
adjusted using the 2 leveling screws on the tooling block
mounting brackets.